User`s manual
30
Input the value you read from FLUEK 45, which is retained to four decimal places.
Press “Clear” key to remove the wrong input value of present digit. Then press
“Enter” key to confirm.
3.2.3 6A Calibration
Press “Enter” key to verify and then you will see the followings:
Input the value you read from FLUEK 45, which is retained to four decimal places.
Press “Clear” key to remove the wrong input value of present digit. Then press
“Enter” key to confirm.
3.2.4 9A Calibration
Press “Enter” key to verify and then the display will show as follows:
Input the value you read from
FLUEK 45, which is retained to four decimal places.
Press “Clear” key to remove the wrong input value of present digit. Then press
“Enter” key to confirm.
3.3 OV Calibration
OV calibration is a kind of automatic calibration without manual operation, which
starts after the voltage calibration is conducted and is shown in the following:
When the calibration is completed, the display will go back to the calibration menu.