User manual

Model: EC-GOLD Dual
ECGOLDDUALUM12.doc Rev: 1.2
17
12. Press <> or <> key to select the #2 Sensor or Exit. If choose the #2 Sensor, the LCD
displays the following:
13. Press <SELECT> Key to the second sensor calibration menu. If the sensor is mA input
sensor, then go to step 14
The calibration procedure for second sensor is the same as the first sensor except for
applying gas to sensor hole (EC2). Follow the steps 3-12, then press <MENU> key to go to
the normal working display.
14. If the sensor is mA input sensor, the calibration of the sensor is done at the remote
sensor (see Remote Sensor Manual). The mA scale is normally factory set, but may be
changed if change needed. Press <SELECT> to change mA settings or press <MENU>
key to bypass the calibration and go back to the Sensor Selection Menu.
15. Press <SELECT> key to set 4 mA, and the LCD displays the following. Press
<SELECT> key to move the cursor and accept the value. Normally, set 4 mA equals 0
ppm. To change the value, press <SELECT> key to move the cursor to proper position
and press <MENU> key to increase up to the right value. Press <SELECT> key to
accept the setting value.
16. Press <SELECT> key to set 20 mA, and the LCD displays the following. Press
<SELECT> key to move the cursor and accept the value. 20 mA is normally set for
maximum range of remote XMTR. To change the value, press <SELECT> key to move
the cursor to proper position and press <MENU> key to increase up to the right value.
Press <SELECT> key to accept the setting value.
17. Then the display goes to the Alarm setting Menu. Press <MENU> key to bypass the
calibration and go back to the Sensor Selection Menu. Repeat Step 12 to Step 16 if
calibrate more sensors.
18. Go back to the normal operating Menu by following the Normal Operating Menu Flow
Chart
THIS COMPLETES THE CALIBRATION PROCEDURE FOR
FULL FEATURE CONTROLLER
Sel Snsr
#1: CO
Sel Snsr
#2: NO2
#2 NO2
Zero
#2 CO
4= 0
#2 CO
mA Scale
#2 CO
20= 200