4.0 CALIBRATION
4.1 Calibration Notes
The instrument was calibrated and tested prior to leaving the factory. Therefore, it is possible to
use the instrument directly out of the box. Under normal conditions, recalibration is
recommended once every 6 months. For a two point calibration, the first point is typically Zero
(clear air); the second point (also called span) is a determined gas concentration value.
If the preset EC-Gold Dual calibration span value is known and you have the same calibration
gas concentration, the entire calibration procedure can be completed without removing the front
cover.
4.2 Calibration Procedures (Full Feature Mode with LCD / KEYPAD)
LCD
SELECT
MENU
EC 1
EC 2
www.arjayeng.com
CANADA LTD.
Calibration
Male, Quick Disconnect
tube insertedinto EC-Gold
Dual Calibration Ports
(EC1 & EC2)
Flowrate Regulator
preset for 0.5 Lpm
Flow On/Off Knob
Cal Gas Canister
6 feet of 5/16" O.D. x 3/16" I.D.
flexible, clear tubing.
EC2
EC1
Calibration Ports
Figure 4.0
1. Under normal operating conditions, the LCD displays the following (depending on
sensors order):
2. Follow the Manu Flow Chart, until the following screen display:
CO 0
NO2
0
#1: CO
Zero