Model: EC-GOLD Dual ECGOLDDUALUM12.doc Rev: 1.2 TABLE OF CONTENTS 1.0 2.0 3.0 4.0 INSTRUMENT OVERVIEW .............................................................................. 3 1.1 Features ................................................................................................ 3 1.2 Description ............................................................................................ 3 1.3 Specifications ...........................................................................
Model: EC-GOLD Dual ECGOLDDUALUM12.doc 1.0 INSTRUMENT OVERVIEW 1.1 Features 1.2 Rev: 1.2 Gas monitor for a large variety of gas types. Can monitor up to 4 different sensor inputs at one time Port 1: on board CO (Carbon Monoxide) Electrochemical Port 2: on board electrochemical for NO2 (Nitrogen Dioxide), CL2 (Chlorine), H2S (Hydrogen Sulfide), NH3 (Ammonia) and others.
ECGOLDDUALUM12.doc Model: EC-GOLD Dual Rev: 1.2 Figure 1.2: EC-Gold Dual can be combined with other EC-Gold Dual or EC-Gold sensors to the ISA Gas Alert Max panel using the cost effective Arjay/Enment DVO signal. Ideal for single or multi-zone installations where the control panel is desirable at each fan location or in central location. Figure 1.3: EC-Gold Dual can be combined with other EC-Gold Dual or EC-Gold sensors to the PG-2000 panel using the RS-485 communication to address each sensor.
Model: EC-GOLD Dual ECGOLDDUALUM12.doc Rev: 1.2 3 Wire Discrete Voltage Output shielded connection To other sensors ZONE 1 ZONE 2 CANADA LTD. ZONE 5 LCD www.arjayeng.com EC 2 BUZZER EC 1 MENU SELECT EC-Gold EC-Gold Dual Zone 5 Relays Zone 1 Relays EC-Gold Dual RLU Application Figure 1.1 3 Wire Discrete Voltage Output shielded connection To other sensors CANADA LTD. LCD www.arjayeng.com EC 2 EC 1 MENU SELECT EC-Gold EC-Gold Dual Relays EC-Gold Dual Gas Alert Application Figure 1.
Model: EC-GOLD Dual 1.3 ECGOLDDUALUM12.doc Rev: 1.2 Specifications OPERATION The EC-Gold Dual is a multi-sensor gas detector. A variety of gases may be sensed when fitted with the appropriate sensors. PUSH-TEST FEATURE On demand using the Push to Test pushbutton.
ECGOLDDUALUM12.doc Model: EC-GOLD Dual 2.0 INSTALLATION 2.1 Mechanical Installation Rev: 1.2 (2) Dual 3/4" & 1/2" Knock-Outs for top entry cabling. 5.66" 6.25" 5.16" Dual 3/4" & 1/2" Knock-Outs for rear entry cabling. DD+ mA/Vout Gnd 24VAC/DC 1 2 3 4 5 CO CO CANADA LTD. 0 0 2 3 C 5 EC 2 4 NC 6 NO 7 C 8 NC 9 EC Sensor Bu zzer 9.61" Status (2 colors) Low Alarm Hi Alarm Hi/Hi Alarm www.arjayeng.com 1 C NC NO Buzzer has built-in ACk.
ECGOLDDUALUM12.doc Model: EC-GOLD Dual Electrical Installation EC-Gold Dual Stand Alone Electrical Installationion Out ! RS-485 Network 24VAC/DC Input 1 2 3 4 5 Use good installation practice! Do not run High Voltage cables in the same conduit as signal wires. JP1 DD+ mA/Vout Gnd 24VAC/DC For Network applications, the 2 wires RS-485 network connects to the PG-2000 Scanner or Central Access Panel.
ECGOLDDUALUM12.doc Model: EC-GOLD Dual Rev: 1.2 EC-Gold Dual and PG-2000 or Central Access Panel Electrical Installationion 4 Wires + Shielded (Shielded connected at this point only) GND 24VAC/VDC R1 R2 DD+ RS-485 Communication Wire (2 Wires) R3 R4 STATUS . 0 ENTER Status (2 colors) 1 NO C 2 NC NO 3 4 C 5 NC 6 NO 7 C 8 NC 9 EC Sensor Buzzer Buzzer has built-in ACk. 1 C 2 NC NO 3 C 5 NC 6 NO 7 C 8 NC 9 Buzzer has built-in ACk.
ECGOLDDUALUM12.doc Model: EC-GOLD Dual Rev: 1.2 EC-Gold Dual and Gas Alert Electrical Installationion The Gas Alert only accommodates a single zone. Sensors per zone: Typically around 20 depending on wires sizes and lengths. SGND connection ( on Gas Alert) NOT made to ED-Gold Dual. N/C 1 2 3 4 5 DD+ mA/Vout Gnd 24VAC/DC DD+ mA/Vout Gnd 24VAC/DC 1 2 3 4 5 C 4 6 NO 7 C 8 NC 9 Status (2 colors) Buzzer has built-in ACk.
ECGOLDDUALUM12.doc Model: EC-GOLD Dual 3.0 Rev: 1.2 STARTUP AND CONFIGURATION User Interface Overview With Cover DD+ mA/Vout Gnd 24VAC/DC 1 2 3 4 5 NO 1 C 2 NC NO 3 C 5 NC 6 NO 7 C 8 NC 9 Low Alarm Hi Alarm Hi/Hi Alarm 0 0 CANADA LTD. LCD www.arjayeng.com EC Sensor Buzzer 4 Buzzer has built-in ACk.
Model: EC-GOLD Dual ECGOLDDUALUM12.doc Rev: 1.2 2. Using the keypad and display (full feature model). The user follows the menus on the display. 3. Using Arjay’s optional 2000-CAL hand held calibrator. The calibrator communicates with the EC-Gold Dual via the RS-485 connection using the modbus protocol. The calibrator’s LCD and keypad offer a more convenient method of setting up and calibrating the EC-Gold Dual unit. 3.2 Startup Power up the unit. The status LED should be green.
Model: EC-GOLD Dual 3.3.1 ECGOLDDUALUM12.doc Normal Operating Menu Flow Chart 13 Rev: 1.
Model: EC-GOLD Dual 3.3.2 ECGOLDDUALUM12.doc Sensors Setup Menu Flow Chart 14 Rev: 1.
ECGOLDDUALUM12.doc Model: EC-GOLD Dual 4.0 CALIBRATION 4.1 Calibration Notes Rev: 1.2 The instrument was calibrated and tested prior to leaving the factory. Therefore, it is possible to use the instrument directly out of the box. Under normal conditions, recalibration is recommended once every 6 months. For a two point calibration, the first point is typically Zero (clear air); the second point (also called span) is a determined gas concentration value.
ECGOLDDUALUM12.doc Model: EC-GOLD Dual Rev: 1.2 3. Zero Calibration only needs to be done when installing a new sensor. Press
ECGOLDDUALUM12.doc Model: EC-GOLD Dual Rev: 1.2 Sel Snsr #1: CO 12. Press <> or <> key to select the #2 Sensor or Exit. If choose the #2 Sensor, the LCD displays the following: Sel Snsr #2: NO2 13. Press
Model: EC-GOLD Dual ECGOLDDUALUM12.doc Rev: 1.2 4.3 Calibration Procedures (Standard Feature Mode Without LCD / KEYPAD) 4.3.1 CALIBRATION NOTES The EC-Gold Dual may be calibrated using a Single Point calibration procedure, by using calibration push button. The Calibration pushbutton may be accessed using a screwdriver or other fine tool – See Figure 4.1 for the location of the access hole.
Model: EC-GOLD Dual ECGOLDDUALUM12.doc Rev: 1.2 2. The gas sensor is mounted directly behind the Calibration Port in the middle of the front plate of the EC-Gold Dual housing [See Fig. 4.1]. Insert the Quick Disconnect tube into the EC-Gold’s Dual Calibration Port. 3. Start gassing the sensor by opening the valve on the canister. The flow rate is preset to 0.5 Lpm (or ~1 SCFH (Standard Cubic Foot per Hour). Wait about 90 seconds. 4.
Model: EC-GOLD Dual 6.0 ECGOLDDUALUM12.doc Rev: 1.2 TROUBLESHOOTING ERROR DESCRIPTION WHAT TO DO 1. No Output and all lights off 2. 4-20mA output does not track the gas concentration. 3. In Discrete Voltage Output mode, the ISA-66RLU or Gas Alert connected to the EC-Gold always shows Sensor Fault 20 Check the power to the unit. The voltage should be 24VAC/DC at the power connector with the positive and negative connected as shown on the connector label.
ECGOLDDUALUM12.doc Model: EC-GOLD Dual 7.0 Rev: 1.2 CONTROLLER SETTINGS SHEET Checked by Model Number Serial Number Hardware Rev. Software Rev. The factory settings column below lists the typical default settings. The user may change these values at any time. If changed, please fill in the USER SETTING column for future reference.
ECGOLDDUALUM12.doc Model: EC-GOLD Dual PARAMETER DESCRIPTION First Cal point SENSOR #1 Enter ppm and corresponding mV e.g. 0ppm= 150mV Second Cal point Enter ppm and corresponding mV e.g. 100ppm=400mV Type or Port. Sensor serial number Gas name Low alarm value SENSOR #2 High alarm value Hi/Hi alarm value Span Value First Cal point Enter ppm and corresponding mV e.g. 0ppm= 150mV Second Cal point Enter ppm and corresponding mV e.g. 100ppm=400mV Type or Port.
MAINTENANCE AND CALIBRATION Your ARJAY / ENMET gas detector is the state of the art in gas detection, but like any other part of your ventilation system, it requires periodic maintenance and calibration. There are various sensor technologies that may be used in your Arjay gas monitor; including solid state MOS, Electrochemical, Infrared and Pellister Pair (Hot wire). These have a normal operating life of 3 to 5 years under ambient conditions.
ARJAY ENGINEERING LTD. GUARANTEE We hereby guarantee this instrument to be free from defects in workmanship and materials, and if found defective in workmanship or materials, upon being returned to our factory, prepaid, within one year from date of purchase, it will be repaired or replaced at factory without charge.