User manual

Model: EC-GOLD
ECGOLDUM17.DOC Rev: 1.7
- 9 -
3.0 STARTUP AND CONFIGURATION
Figure 3.0
3.1 NOTES ON THE USER INTERFACE OPTIONS
The EC-Gold may be configured and calibrated using either of 2 methods:
1. Using the integral DIP switch and Calibration pushbutton. The EC-Gold front cover must be
removed to access the DIP switch. For calibration only, the pushbutton is accessible without
removing the front cover.
2. Using Arjay’s optional 2000-CAL hand held calibrator. The calibrator communicates with the EC-
Gold via the RS-485 connection using the modbus protocol. The calibrator’s LCD and keypad
offer a more convenient method of setting up and calibrating the EC-Gold unit.
3.2 STARTUP
Power up the unit. The status LED should be green. The High and Low alarm LED’s should go off in
less than a minute.
The unit is normally pre-configured and calibrated at the factory so field setup is not required on
startup.
To check calibration, or to re-enter / modify settings, follow the procedures described below. The method
used will depend on if the integral user interface is used (DIP switch / pushbutton) or if the Arjay/Enmet
USRINT.DSF
12 3 4567
Micro
Calibration
pushbutton
SENSOR BOARD
+24V
Gnd
mA / Vout
D+
D-
RS-485
Push to Test
pushbutton
USER INTERFACE OVERVIEW
12 34567
ON
7 POSITION DIP SWITCH
High Alarm
Low Alarm
Status (2 color)
STATUS LED COLOR
Green
Green blinking (ON and OFF)
Orange (Red & Green both ON)
Red (On for 2 seconds only)
Red (Steady ON)
Instrument ok, no faults. Will remain green for High or Low Alarms
In Configuration entry mode or while waiting for 2nd Calibration point.
Calibration request acknowledge - while cal pushbutton pressed
Calibration error or user error.
Instrument failure / Sensor Fault condition
CO sensor shown for example
Push to Test only available
when used with CO sensor
MEANING
1
2
3
4
5
TB1
The DIP switch and the
Calibration pushbutton are
used to configure / calibrate
the unit.
Calibration
pushbutton
Push to Test
pushbutton