MODEL 9830-OWS OIL/WATER INTERFACE LEVEL SWITCH User Manual Arjay Engineering Ltd. Oakville (Toronto), Canada, L6H 6C9 Tel . Fax. ++1 (905) 829-2418 ++1 (905) 829-4701 North America 1-800-387-9487 www.arjayeng.com arjay@arjayeng.
9820 | 9830-OWS Oil / Water Separator Level Alarm Reliable interface monitoring of oil/water separators, sump pits and containments for pump control and alarm Over 30 years of Arjay’s field proven HF capacitance technology has been applied to the 9820/9830-OWS oil water separator alarms. The sensing probe continuously monitors for oil to control pumps, operate valves, or activate alarms.
9820 | 9830-OWS The 9820/9830-OWS sensing probe monitors the capacitance field around the active probe tip. As the volume of separated oil increases in the separator or is drawn down toward the probe tip, the probe capacitance changes. This change is used to activate the relay for alarm interface and control.
MODEL DOCUMENT TYPE DOCUMENT FILE NAME 9830-OWS USER MANUAL 9830OWSUM14.DOC 1.4 REV. CREATE DATE REV. DATE PRINT DATE 04/08/2003 12/22/2009 10:07 AM 12/22/2009 10:07:00 AM TABLE OF CONTENTS 1.0 2.0 3.0 4.0 INSTRUMENT OVERVIEW............................................................................................ 3 1.1 FEATURES ....................................................................................................... 3 1.2 DESCRIPTION ........................................
MODEL DOCUMENT TYPE DOCUMENT FILE NAME 9830-OWS USER MANUAL 9830OWSUM14.DOC 1.4 CREATE DATE REV. DATE PRINT DATE 04/08/2003 12/22/2009 10:07 AM 12/22/2009 10:07:00 AM 1.0 INSTRUMENT OVERVIEW 1.1 FEATURES REV. Push-button calibration RF Technology Double pole Double Throw (DPDT) 5A relay contacts 0 – 20 second time delay Adjustable sensitivity High or Low relay action No moving parts For use with any Arjay Level Probes 1.
MODEL DOCUMENT TYPE DOCUMENT FILE NAME 9830-OWS USER MANUAL 9830OWSUM14.DOC 1.4 CREATE DATE REV. DATE PRINT DATE 04/08/2003 12/22/2009 10:07 AM 12/22/2009 10:07:00 AM 9830-OWS OIL / WATER SEPARATOR APPLICATION 2 WIRE (SHIELDED) INTRINSIC SAFETY BARRIER (OPTIONAL) 9830-OWS Manway Arjay Level Probe Earth Line Oil Pump Overflow Liquid Level Sensing portion of probe is below pump suction OIL WATER SEPARATOR 9830 OWS BLOCK.DSF Figure 1.
MODEL DOCUMENT TYPE DOCUMENT FILE NAME 9830-OWS USER MANUAL 9830OWSUM14.DOC 1.4 REV. CREATE DATE REV. DATE PRINT DATE 04/08/2003 12/22/2009 10:07 AM 12/22/2009 10:07:00 AM OPERATION The unit uses an RF Capacitance measurement technique for high resolution measurements. When the presence of oil displaces the water the level probe senses the decrease in capacitance and the control unit will indicate an alarm condition and change the state of it’s dry relay contacts.
MODEL DOCUMENT TYPE DOCUMENT FILE NAME 9830-OWS USER MANUAL 9830OWSUM14.DOC 1.4 CREATE DATE REV. DATE PRINT DATE 04/08/2003 12/22/2009 10:07 AM 12/22/2009 10:07:00 AM 2.0 REV. INSTALLATION NOTE: If any damage to the instrument is found, please notify an Arjay Engineering representative as soon as possible prior to installation. Choose the mounting location in accordance with good instrument practice. Extremes of ambient temperature and vibration should be avoided. 2.
MODEL DOCUMENT TYPE DOCUMENT FILE NAME 9830-OWS USER MANUAL 9830OWSUM14.DOC 1.4 CREATE DATE REV. DATE PRINT DATE 04/08/2003 12/22/2009 10:07 AM 12/22/2009 10:07:00 AM THREADED ENTRY FLANGED ENTRY CONCENTRIC SHIELD ENTRY Use wrench on Lower Hex ONLY 2" Entry Typical 1- For threaded and flanged entry types, measurement sensitivity is increased by reducing the probe to wall distance. 2- There should be good electrical conductivity between the tank wall and the transmitter enclosure.
MODEL DOCUMENT TYPE DOCUMENT FILE NAME 9830-OWS USER MANUAL 9830OWSUM14.DOC 1.4 CREATE DATE REV. DATE PRINT DATE 04/08/2003 12/22/2009 10:07 AM 12/22/2009 10:07:00 AM 2.4 REV. ELECTRICAL INSTALLATION ALARM STATUS Single DPDT relay with dry contacts rated at 5A @ 250VAC or 30VDC for resistive loads.
MODEL DOCUMENT TYPE DOCUMENT FILE NAME 9830-OWS USER MANUAL 9830OWSUM14.DOC 1.4 CREATE DATE REV. DATE PRINT DATE 04/08/2003 12/22/2009 10:07 AM 12/22/2009 10:07:00 AM 3.0 REV. STARTUP AND CALIBRATION ALARM STATUS Calibrate Pushbutton CALIBRATE DIELECTRIC Normally set for HIGH HIGH 6 5 4 Normally set for 3 6 5 4 8 7 ALARM DELAY 9 0 1 2 3 8 7 SENSITIVITY 9 0 1 2 3 LOW Sets Failsafe of relay contacts RELAY LOW Alarm delay: 0 - 20sec. Each switch inc. = 2sec except 6=15 and 7=20.
MODEL DOCUMENT TYPE DOCUMENT FILE NAME 9830-OWS USER MANUAL 9830OWSUM14.DOC 1.4 CREATE DATE REV. DATE PRINT DATE 04/08/2003 12/22/2009 10:07 AM 12/22/2009 10:07:00 AM 3.2.2 REV. Alarm Time Delay This is set via the Alarm Delay switch. An alarm based on the Relay Action switch must be present for at least the time delay value for the relay to switch to the alarm state. There is no delay when switching off the alarm. The delay range is 0 – 20 seconds.
MODEL DOCUMENT TYPE DOCUMENT FILE NAME 9830-OWS USER MANUAL 9830OWSUM14.DOC 1.4 REV. CREATE DATE REV. DATE PRINT DATE 04/08/2003 12/22/2009 10:07 AM 12/22/2009 10:07:00 AM Make sure the Oil/water Separator is filled with water to capacity or the Sump level above the active portion of the probe. With the active sensing portion of the probe submerged in water, push the CALIBRATE BUTTON. The status LED will turn red momentarily. Release the button.
MODEL DOCUMENT TYPE DOCUMENT FILE NAME 9830-OWS USER MANUAL 9830OWSUM14.DOC 1.4 CREATE DATE REV. DATE PRINT DATE 04/08/2003 12/22/2009 10:07 AM 12/22/2009 10:07:00 AM 3. False alarms REV. Add some time delay to unit. If coax cable is used from sensors to pulse card (PMC 2000), make sure it is not coiled (may cause an increase in inductance).