Chapter 5: Sorting 167
NOTE Before beginning these procedures, make sure the stream is stable and the
Sweet Spot is on.
Setting Up the Experiment
The steps in this section show you how to set the drop delay using the Accudrop
experiment template. Because no data is recorded, the experiment can be reused
as often as you like.
1 Create an experiment from the Accudrop Drop Delay template.
Select Experiment > New Experiment. Select the Accudrop Drop Delay
experiment and click OK.
2 Expand Specimen_001 and Tube_001.
3 Set the current tube pointer to Tube_001.
4 Open the sort layout by double-clicking it.