User Manual
7
Cleaning and Care
The Omega Pipettor can be cleaned with most common laboratory disinfectants, including
isopropanol. Never use aggressive organic solvents.
The nosepiece can be autoclaved at 121°C, with the exception of the filter. Follow the
procedure under “Replacing the Filter” to remove the filter prior to sterilization.
Trouble Shooting
Problem Possible Cause Remedy
Slow Aspiration Filter Blocked Replace Filter
Slow Aspiration Battery Low Recharge Battery
Pipette Drips Pipette not secure in holder Re-insert Pipette into Pipettor
Pipette Drips Pipettor Mount Damaged or Worn Replace Pipettor Mount
Note that there are no user serviceable parts inside of the unit.
Replacement Parts Ordering Information
Omega™ Pipettor
Part No. Description
P5000 Omega Pipettor, Grey
P5015 Omega Pipettor, Green
P5016 Omega Pipettor, Red
P5017 Omega Pipettor, Blue
P5018 Omega Pipettor, Clear
P5019 Omega Pipettor, Purple
P1020
Silicone Pipette Holder
153 015 Sterile Filters (0.45µm) 10/PK
153 014 Non Sterile Filters (0.45µm) 10/PK
153 017 Barrier Filters (0.45 / 0.8µm) 10/PK
P3225 Pipettor Mount
P5025 Nosepiece – Red
P5070 Nosepiece – Grey
P3010
Battery Charger
P1005 Pipettor Wall Mount
P5022 Personalized Name Card