User manual

48 FastSCAN
Figure 30: Three of the four calibration spot clusters are visible (for the Scorpion
TM
)
and the Calibration Correction dialog box.
Notes:
You must collect the 100 mm data first.
Windows 2000 or XP users must have Power User privileges or greater.
If desired the most recent sweep (or the point in this case) can be removed with
a double click of the trigger, or all data can be removed with a triple click. These
functions are switched on by default—see Section 3.4.
To help reduce artifacts the data is restricted to points within the vicinity of the
Calibration Target; it will not matter if you place the Transmitter and Target on a
light-colored surface.
Should the Transmitter already be bolted to a surface then the Calibration Target
may be placed on the opposite face. It is imperative that the target is mounted
securely.
To remove all Calibration Corrections made to date click the Revert to Factory
Calibration button.