Figure 30: Three of the four calibration spot clusters are visible (for the Scorpion
TM
)
and the Calibration Correction dialog box.
Notes:
• You must collect the 100 mm data first.
• Windows 2000 or XP users must have Power User privileges or greater.
• If desired the most recent sweep (or the point in this case) can be removed with
a double click of the trigger, or all data can be removed with a triple click. These
functions are switched on by default—see Section 3.4.
• To help reduce artifacts the data is restricted to points within the vicinity of the
Calibration Target; it will not matter if you place the Transmitter and Target on a
light-colored surface.
• Should the Transmitter already be bolted to a surface then the Calibration Target
may be placed on the opposite face. It is imperative that the target is mounted
securely.
• To remove all Calibration Corrections made to date click the Revert to Factory
Calibration button.