Owner manual

The P60C-5 pH and the R60C-5
ORP probes are the latest additions
to the legendary Aquametrix differ-
ential probe family. It lls the gap
for a 1” xed insertion, MNPT probe
and uses the same electrodes and
salt bridge as the 1variable inser-
tion P60C-6 probe.
All of the P60 and P65 probes are
dependable industrial grade sensors
designed to provide accurate mea-
surement and longer service life un-
der the most demanding conditions.
Some features of these probes
include: differential measurement
technology, replaceable salt bridge
and encapsulated preamp.
The P/R60C-5 incorporates all the
benets of differential measure-
ment found in other models of the
60 series, eld proven in thousands
of installations. This technique uses
two glass electrodes to make the
measurement differentially with re-
spect to a third metal electrode. The
domed glass process electrode is
specially designed for tough ap-
plications. The second electrode is
immersed in a pH 7 buffer encapsu-
lated in the probe.
This second electrode is protected
from the process by a double junc-
tion replaceable salt bridge. The re-
sulting true differential measurement
has several advantages over con-
ventional probes: ground loop prob-
lems are virtually eliminated, and the
salt bridge is easily replaced.
If the internal solution becomes con-
taminated, the probe can be rejuve-
nated at modest cost by replacing
the salt bridge and reference solu-
tion. Automatic temperature com-
pensation is accomplished through
the use of a thermistor at the tip of
the probe. This thermistor place-
ment provides rapid response for
process temperature variations.
The encapsulated preamplier pro-
vides an output signal which can
be transmitted 3000 feet over in-
expensive cable. Another version
encapsulates a blind 4-20 mA two
wire transmitter which can transmit
a virtually unlimited distance over a
twisted pair cable. (See P/R65 data
sheet).
Features
Replaceable Salt Bridge
Low Maintenance Cost
Field-proven
Encapsulated Preamplier
Transmits up to 3000 ft.
Options
4-20 mA two-wire blind transmit-
ter version available (model P/R65)
ORP available with gold electrode
Aquametrix Model P/R60C-5
Differential pH or ORP Probes
by
Water Analytics
Aqua
Metrix
100 School Street
Andover, MA 01810
978-749-9949
Toll free - 855-747-7623
www.WaterAnalytics.net
Applications
Process Control
Industrial and Municipal Waste
Treatment and Neutralization
Fume Scrubbers
Plating
Circuit Board Manufacturing
Food and Beverage
Chemical Processing
Pulp and Paper
Mining
Power Generation
Description
by
Water Analytics
Aqua
Metrix
New

Summary of content (2 pages)