User Manual Owner manual

2
1. Overview
The AM-TBR process turbidimeter allows for the measurement of the turbidity of process water
on-line. The white light AM-TBR has been designed to meet the design criteria specified by the
US EPA 180.1 on turbidity measurement. The infrared AM-TBR was designed to meet the
design criteria specified in ISO 7027 and DIN 27027 for the measurement of the turbidity of a
sample. All models have long life lamps.
Some models have ultrasonic cleaning. Refer to 8.2 for more information.
A pressure regulator on the incoming line is standard on all AM-TBR models and will
reduce pressures up to 1380kPa (200 PSI) down to (104kPa) 15 PSI.
1.1. The AM-TBR Series
The AM-TBR series instruments have two light source options available, and two range options
available. The only other options are the built-in ultrasonic cleaning, and the flow alarm relay.
All units have user selectable 4-20 mA or Modbus over RS-485 outputs. Refer to the table below
to determine which factory installed options are available. Add the postfix A to the model
number for the alarm option.
Model Number
Illumination
Type
Ultrasonic
Cleaning
AM-TBR2 WL
White Light
No
AM-TBR2 IR
Infrared
No
AM-TBR3 WL
White Light
Yes
AM-TBR3 IR
Infrared
Yes
AM-TBR4 WL
White Light
Yes
AM-TBR4 IR
Infrared
Yes
1.2. Unpacking and Inspection of the Instrument and Accessories
The table below indicates the items in the turbidimeter shipment.
Item
Quantity
AM-TBR Turbidimeter c/w Field Terminal Box & Flow Through Assembly
1
Instruction Manual
1
Desiccant Pack
1
Cuvette (Single Pack)
1*
Tubing Kit: 1-shutoff clamp
1-backpressure valve
2-connecting tubing with fittings for flow through assembly
1-drain vent screw (used in pressurized systems)
1