the instrument, of the new desiccant it will be necessary to reset the instrument by disconnecting
the sensor interconnect cable for 2 seconds and then reconnecting it.
Note: Once the bag is opened, install the desiccant pouch immediately to prevent premature
degradation of the desiccant.
10.3. Replacing the Source Lamp
The source lamps in the AM-TBRs are designed for long life. The IR lamp is rated for 10 years
and the white light version is rated for 7 years. If the lamp should need replacement, we
recommend calling WA Service Department for assistance.