OPERATING INSTRUCTION MANUAL MODEL 2200P PH ANALYZER REV. 12 Water Analytics, Inc.
www.Aquametrix.
MODEL 2200P pH CONTROLLER INSTRUCTION MANUAL 1.0 GENERAL INFORMATION The AquaMetrix Model 2200P pH analyzer is a versatile industrial microprocessor based instrument. Setpoints and outputs are programmed through the menu with push buttons on the face of the instrument. Calibration is achieved from the front panel menu. The instrument may be used in conjunction with AquaMetrix 5-wire differential probes or with any conventional combination probes.
2.0 SPECIFICATIONS DISPLAY: 4 x 7 segment ½” LED Display MEASURING RANGES: pH: 0.01 to 14.00 pH Temperature: 0°C to 100°C (32°F to 212°F) POWER REQUIREMENTS: Standard: 98-132 Vac, 50/60 Hz (less than 20 VA) Optional: 196-264 Vac, 50/60 Hz (less than 20 VA) 23-26Vdc (nominal current 150mA) AMBIENT CONDITIONS: -30 to 50°C (-22 to 122°F) 0 to 90% R.H.
3.0 INSTALLATION 3.1 Location 3.1.1 Install the instrument within 3000 feet of where the AquaMetrix differential probe (P60 Series) is installed. If a conventional combination probe (P500 Series) is used the instrument must be within 10 feet of the probe for direct connection. An AquaMetrix 101 Series preamplifier may be used to extend this distance to 3000 feet. 3.1.
3.4 Electrical Connections The 2200P is available in 3 different power options, 120Vac, 240Vac, and 24Vdc. Refer to following sections for connecting the input power. 120Vac and 240Vac 3.4.1 To access the terminal strips open the door of the instrument and then unscrew the captive retaining screw near the upper right hand corner of the panel. Now swing open the panel to reveal the terminal strip on the power supply circuit board and the smaller terminal strip on the back of the main circuit board.
24Vdc 3.4.3 To access the terminal strips open the door of the instrument and then unscrew the captive retaining screw near the upper right hand corner of the panel. Now swing open the panel to reveal the terminal strip on the power supply circuit board and the smaller terminal strip on the back of the main circuit board. 3.4.4 The terminal strip on the power supply board at the back of the instrument is labeled for input power, relay outputs and analog outputs.
3.5 Probe Connections: DIFFERENTIAL PROBE (P60 SERIES) 3.5.1 Ensure that both J42 jumpers are in the up position, leaving the pins nearest to the bottom exposed. Connect the 6 wires from the probe to the TB2 terminal strip on the main circuit board. Match the colors of the probe wires to the terminals marked on the circuit board. (Refer to Fig. 3) 3.6 Sensor Connections; Combination Probe (P500K Series) With Temperature Compensation 3.6.
b) Connect the active electrode (ring terminal) to the terminal post on the swingout board and connect the shield to the SH terminal of the terminal strip TB2. c) Connect the temperature sensor wires to YL and BK terminals of TB2 disregarding the colors. 3.7 Sensor Connections; Combination Probe (P500 Series) Without Temperature Compensation 3.7.1 a) Both J42 jumpers located on the swing-out board must be in the down position leaving the pin nearest to the top exposed. (Refer to Fig.
be transmitted up to 3000 feet, the same as the differential (P60) probe. When the pre-amplifier is used the controller must be configured, as if you were using a differential (P60) probe. (Refer to FIG. 6) 4.0 Description of Functions 4.1 Overview Model 2200P 4.1.1 The Model 2200P is a microprocessor operated pH analyzer designed for industrial applications. It is compatible with a wide range of AquaMetrix pH probes and offers several measurement ranges.
performs an extensive system of checking values and parameters. All of the operating parameters are stored in non-volatile memory, without the need of a battery. 4.1.4 The Model 2200P operates like a normal analog converter with a number of additional functions made possible by the microprocessor in the instrument.
the Normally Open contact. In the event of power loss to the unit, the relay will deenergize and the device connected to the Normally closed contact will be turned on. This option is selected with the DIP switches on the swing out board. See Section 4.7. 4.4.5 An option provides cycling for Relay A and Relay B to obtain closer process control. When the process is between the relay on and relay off setpoints the relay will cycle on and off, based on the cycle times selected by the user.
4.6.4 When in the menu mode, the display initially shows the current value of the parameters, such as the setpoint of the control relay, while putting all of the outputs on hold. The two arrow buttons are used to adjust the display value up or down. To accept the new value press ENTER twice. While the value on the display is being changed, the relay outputs and the analog outputs remain on hold.
Bank S2 Dip Switch 1 2 3 4 5 6 7 8 Description Fail-Safe mode for Relay B Direction of Control for Relay B Cycling Control for Relay A Cycling Control for Relay B Reserved Reserved Reserved Reserved ON Enabled Falling Enabled Enabled OFF Disabled Rising Disabled Disabled Note: Items in “BOLD” are the factory settings that the unit will be shipped with. 4.8 Output Hold 4.8.
4.13 Watchdog Timer and Self Diagnostics 4.13.1 The 2200P continuously monitors the condition of all key components of the measuring system to ensure that the measurements are reliable. Invalid entries and memory loss are indicated on the panel. See Section 8.0. 4.14 Cycle Feature 4.14.1 With this feature the output of Relay A and/or Relay B can be cycled. The ON time and the OFF time are independently adjustable. This feature can help eliminate the problem of overshooting. Saving expensive chemicals. 5.
7 pH as your first value, and the other as the second. If you are calibrating with 4 pH, and 10 pH, use 4 pH as the first and 10 pH as the second. To calibrate the 2200P in Manual Calibration Mode proceed as follows: a) Enter the menu by pressing MENU. Proceed by pressing MENU until the CALIBRATION LED illuminates. b) Remove the protective cap from the probe, rinse it in clean water, and place it in the first buffer solution.
receiving an unstable signal from the probe. When the display is steady, Press ENTER. The display will flash until ENTER is pressed again to confirm the value. c) Rinse the probe in the clean water, and place it in the second buffer solution. Again the instrument will recognize the buffer and will show the value on the display. When the display is steady, Press ENTER. The display will flash until ENTER is pressed again to confirm the value. d) The system is now calibrated.
the display read the desired value. Press ENTER. The display will flash until ENTER is pressed again to confirm the value. b) Press RUN, to place the instrument on line, or press MENU for another menu selection. 5.6 Control Relay A Relay Off Setpoint 5.6.1 If you have configured the relay to control decreasing pH, the relay off setpoint will be higher than the relay on setpoint. i.e. the auxiliary device connected to the relay will increase the pH.
• Set Control Relay B Relay Off Setpoint at 7.50 pH With the above configuration Relay A will control the addition of caustic. The caustic pump will turn on if the pH falls to 6.00 pH and will stay on until the pH rises above 6.50. Relay B will control the acid pump. The acid pump will turn on if the pH rises to 8.00 pH and will stay on until the pH falls to 7.50. Therefore the pH will always remain between 6.00 pH and 8.00pH 5.9 Cycle Feature for Relays A and B 5.9.
as follows. a) Press MENU to enter the menu, proceed by pressing MENU until the ALARM HIGH LED is illuminated. Use the arrow keys make the display read the desired value. Press ENTER. The display will flash until ENTER is pressed again to confirm the value. b) Press RUN to place the instrument on line, or press MENU, for another menu selection. 5.11 Alarm Low 5.11.1 To set the ALARM LOW proceed as follows: a) Press MENU to enter the menu, proceed by pressing MENU until the ALARM LOW LED is illuminated.
5.13.2 The 0-5 Vdc and 0-1 mA analog outputs can be selected to track the process temperature by simply placing DIP switch No. 8 of Bank S1. in the OFF position. The temperature span of the output is set to the utility menu. See Section 7.4. 5.14 Status 5.14.1 The 2200P continuously checks the integrity of all stored data and monitors the condition of the measuring system. If a fault is detected, the STATUS LED above the display will change from green to red.
6.1.3 If the system can no longer be calibrated, that is, if you cannot make the instrument read the high or low buffer value, then the probe may need to be replaced. 6.2 Calibration 6.2.1 The calibration procedures outlined in Sections 5.2 and 5.3 are recommended. However, it is possible to calibrate with one point only or to any known value if you prefer but only if a successful 2 point calibration with 2 buffers has been performed previously. 6.2.
7.2 Access to Utility Menu 7.2.1 The Utility Menu is protected by password. To access the Utility Menu press and hold both RUN and ENTER for five seconds until the PASSWORD LED illuminates. The green STATUS light will flash. Now with the arrow keys make the display read the password, 7. Press ENTER. 7.2.2 The Utility Menu is entirely separate from the Operation Menu but uses the same menu LED’s. Press MENU to step through the menu items.
Range 7 pH 4 pH 10 pH 0 to 14 pH 2 to 12 pH 0 to 10 pH 4 to 10 pH 2.5 2.5 3.5 2.5 1.43 1.0 -0.0 3.57 4.0 5.0 5.0 f) After returning to on-line, recalibrate as described in Section 5.2. 7.4 Temperature Calibration 7.4.1 The temperature channel was calibrated at the factory. It is a single point calibration at 25°C. If needed, this calibration can be repeated as follows: a) Disconnect the temperature sensor wires from terminals BK and YL on TB2. b) Enter the Utility Menu, as described in 7.2.1.
7.6 Adjust 0-1 mA / 0-5 Vdc Output 7.6.1 It may be desirable to fine tune the 0-1 mA / 0-5 Vdc output to take into account the characteristics of your particular loop. The following method involves a high and low calibration and requires the use of a digital multi-meter. (DMM). Proceed as follows ignoring the instrument display: a) Turn off the power to the instrument. Connect your DMM in parallel with the 0-5 Vdc output terminals on the power supply board. b) Turn on the power.
8.1.1 The 2200P has diagnostic features which alerts the operator to invalid entries and memory loss. Invalid entries are indicated by the flashing of the appropriate menu LED. The flashing will commence after RUN is pressed and will continue until the errors are corrected. Memory loss is indicated by the flashing of TEST and by the alarm relay if enabled by DIP Switch No. 5 of Bank S1. See Section 4.7.3. 8.2 Invalid Calibration 4.8.
these steps do not solve the problem it may be necessary to replace the power supply board. See Section 9.4 below. 9.2.3 Should your process control element e.g. pump, valve or alarm not be activating, check that the correct power is supplied to the "floating” relay contacts, jumped over from the line voltage terminals or perhaps from a separate breaker or DC supply. Use the TEST menu feature to simulate the input values to observe the switching points.
case of a memory loss problem, when so advised by AquaMetrix service support. 9.3.3 Before performing the "ESCAPE" procedure it is important to know that this procedure provides the option to erase all programmed values and replace them with the factory set default values. This is true for the control relay on and relay off setpoints, the alarm settings and the analog outputs scaling. Also, the temperature and the pH calibration points will be set to their initial values.
b) The power supply circuit board is fastened to the back of the enclosure by four screws, remove the screws and the board is free. Reverse the procedure to mount a replacement board. c) The microprocessor circuit board is located on the swing-out assembly behind the door. Swing the assembly out, lift it up to unseat the lower hinge pin (the upper hinge pin is spring loaded.) The assembly is now free. The circuit board is fastened to the front panel by five screws.
STATEMENTS OF CONFORMITY FROM THE MANUFACTURER WARNING: This equipment generates, uses, and can radiate radio frequency energy and if not installed and used in accordance with the instructions manual, may cause interference to radio communications. It has been tested and found to comply with the limits for a Class A computing device pursuant to Subpart J of Part 15 of FCC Rules, which are designed to provide reasonable protection against such interference when operated in a commercial environment.