Operation 4-25
4.3.1.3 Measurement Setup Based On Existing Setup
This measurement setup assumes there are existing relevant setup files for the B Coil, H Coil, and Gaussmeter
probe you are about to use. This procedure opens an existing setup
1. Apply power to the 7800 Series Hysteresisgraph System and launch the Hysteresisgraph software.
2. In the Security menu, select Change User Mode. Enter the password to switch to Supervisor Mode.
3. In the File menu, select Open Setup. Locate, highlight, then open the desired .SET file.
4. In the Setup menu, select System Setup. Change the appropriate fields of information detailed in
Paragraph 4.1.3.1.
5. In the Setup menu, select Calibration Setup. Change the appropriate fields of information detailed in
Paragraph 4.1.3.2.
6. In the Setup menu, select Unit Select. Change the appropriate fields of information detailed in Paragraph
4.1.3.4.
7. In the File menu, select Save Setup. Setups are saved as files with the .SET extension. The file name
should end with the 10-digit (maximum) serial number. For example, *_HYSBH00002.SET.
4.3.2 Operator Mode
In the Operator Mode, the Setup, Sensing, and Data Analysis menus are grayed out, meaning their functions are
not assessable. The Save Setup selection in the File Menu is also grayed out.
1. Run the appropriate AutoCalibration Mode. Follow instructions in Paragraphs 4.2.1 thru 4.2.3 as
appropriate.
2. For single measurements, follow the Run One Sample instructions in Paragraph 4.1.5.1.
3. For multiple measurement using the same system setup, follow the Run Sequence instructions in Paragraph
4.1.5.2.