User`s manual

7800 Series Hysteresisgraph System User's Manual
Operation 4-23
Mode 3 Calibration (Continued)
8. While holding the reference standard in the center, place the B coil inside the electromagnet gap, then click
the OK button.
9. Remove the reference/coil from the electromagnet, then click the OK button.
10. After waiting for the field to settle, the software will display the drift. Click the Accept button to proceed
with the Mode 3 calibration. Press the Repeat B Calibration button to return to Step 7.