4-22 Operation
Mode 3 Calibration (Continued)
5. After waiting for the H field to settle, the software will display the accumulated drift. Click the Accept
button to proceed with the Mode 3 calibration.
6. Check the power supply to ensure the ON and REMOTE indicators are illuminated. Click the OK button
to turn power supply control over to the computer.
7. Place the sensing coil at a point of minimum flux (hold in a vertical orientation away from the
electromagnet), then click the OK button.