Operation 4-15
Data Processing (Continued)
OFF No QC checking occurs. The Main Screen will not display Pass/fail or error data on this item.
Lo/Pass QC checking is turned ON. Testing will result in either a Lo test result or a Pass test result. A
lower test limit must be assigned.
Hi/Pass QC checking is turned ON. Testing will result in either a Hi test result or a Pass test result. A
higher test limit must be assigned.
Lo/Pass/Hi QC checking is turned ON. Testing will result in either a Hi test result , a Lo test result, or a Pass
test result. A lower test limit and a higher test limit must be assigned.
Br Remanence Check to display the Br – Remanence value onscreen
Hc Normal Coercivity Check to display the Hc – Normal Coercivity/Coercive Field Strength of
Magnetic Flux Density, onscreen
Hci – Intrinsic Coercivity Check to display the Hci – Intrinsic Coercivity/Coercive Field Strength of
Magnetic Polarization, onscreen
BH(max) – Maximum energy product Check to display the BH(max) – Maximum energy product,
onscreen
Drift in B Channel Check to display the Accumulated Drift in the B channel fluxmeter at the end of the
measurement process.
Drift in H Channel Check to display the Accumulated Drift in the H channel fluxmeter at the end of the
measurement process.
Summed (B-H) Drift Check to display the difference in the drift accumulated by the B and H integrators –
see Drift B or Drift H.