User`s manual
7800 Series Hysteresisgraph System User's Manual 
4-10  Operation 
4.1.4  SENSING MENU (Supervisor Mode Only) 
The Sensing Menu consists of B, H Coils in Paragraph 4.1.4.1, Hall Probe in Paragraph 4.1.4.2, and Write to 
Coil EEPROM in Paragraph 4.1.4.3. These menu selections are only available when the Hysteresisgraph 
software is operating in Supervisor Mode. 
4.1.4.1  B, H Coils 
The B Coil Setup window is shown below. The H Coil Setup window requires similar information. 
Coil Input Terminal  Choose 10 kΩ or 100 kΩ for the input resistance of the fluxmeter. For most Lake 
Shore coils, this setting should be 100 kΩ. This selects the input resistor to the 
integration stage of the integrator. Choosing a value of 10 kΩ instead of 100 kΩ will 
increase the system gain by a factor of 10 and increase the fluxmeter sensitivity to 
drift by a factor of 10. If this amount of gain is not required by your measurement, it 
should not be used. 
Number of Turns  Enter the number of turns marked on your sensing coil. 
Average Area/Turns  Enter the average area of the entire coil structure (average area = A·N/turns). 
Area Turns  Enter the total area·turns (A·N) for your sensing coil. 
Coil Resistance  Enter the DC resistance of the attached sensing coil. 
4.1.4.2 Hall Probe 
Serial Number  Enter the serial number of your Hall probe. 
Date  Enter the date of manufacture of your Hall probe 
Nominal Size  Enter the nominal size of your Hall probe. 
Comment  Enter any additional information concerning your Hall probe. 