Instruction manual
KT series chillers – Installation and Operation manual 
Issue 7.0 Page 9
3.0   Operation 
KTR models only 
Connect the mains cable and switch the unit on. The neon will illuminate, and the chiller 
runs automatically. KTR chillers have been configured to provide temperature stability to 
±0.2°C, and the actual temperature achieved will be in the range 12°C-20°C, depending on 
the load and ambient temperature. This model is designed to provide a reliable and 
inexpensive alternative to tap water. 
The pump in the standard version KTR6*** may require priming, simply achieved by 
switching the unit on and off repeatedly until no bubbles are seen in the tank. 
KTC models only 
Connect the mains cable and switch the unit on, and the temperature display will 
illuminate. KTC chillers have been configured to provide temperature stability to ±0.2°C, 
and the actual temperature achieved will be in the range 12°C-20°C, depending on the load 
and ambient temperature. This model is designed to provide a reliable and inexpensive 
alternative to tap water. 
The KTC includes digital coolant temperature display and high/low temperature and low 
fluid level alarms. The single neon and                icon provide visual alert to the 
alarm conditions: 
High temperature alarm: Preset at 15°C – neon only 
 Preset at 25°C – neon, buzzer* and HI warning display 
Low temperature alarm: Preset at 8°C – buzzer* and LO warning display 
Low fluid alarm:  Neon and buzzer* only 
* The buzzer can be muted by pressing the MUTE button on the controller. 
No adjustments can be made to the factory-set cooling parameters. 
Operator 
section 










