Instruction manual
KT series chillers – Installation and Operation manual 
Issue 7.0 Page 11
Fluid flow and pressure 
KT chillers fitted with a 2 or 3 litre per minute pump are capable of producing up to 150 
psi, because of the high pressure volumetric pump. These units are normally supplied 
factory set to run at an operating pressure of 35 psi, which is suitable for most applications. 
The pressure setting is controlled by an internal bypass valve, which will open above the 
preset level to protect your application from higher pressures. 
It is possible for customers to change the operating pressure of all K series chillers, except 
KT chillers fitted with a 6 litre/minute pump
1
. For details of this procedure, refer to the 
technician section. 
4.0   Maintenance requirements, Operator 
Caution: Failure to carry out service at the specified intervals may permanently damage 
your equipment. For maintenance and service requirements for technicians, and for 
troubleshooting, refer to the technician section. 
Exterior cleaning: Wipe up any spillages using a damp cloth only. Do not use detergents. 
Interval Actions 
Weekly Check fluid level 
Check external connections for leaks 
Annually Change the fluid 
1
 The 6 litre/minute pump is specially designed for low pressure applications, and the flow rate and pressure 
settings cannot be changed on these units. 
i
Operator 
section 