The AV760 is a precision instrument and must be
maintained to ensure proper functions. If the AV760
sensors get dirty and no longer read correctly, the
Sensor Housing can be removed to enable easy
cleaning of the sensors.
1. Remove the tting, o-ring, and the oil catch lter
from the sensor housing.
WARNING: Do NOT remove the protective screen
covering the sensor.
2. Apply Isopropyl Alcohol (70% or higher) to the
sensor ports and flush the sensor ports for 5
minutes (page 9).
WARNING: Do NOT touch the sensors or insert
any objects into the sensor ports. Doing so may
permanently damage the sensors.
3. Drain the uid out of the sensor and apply a short
low pressure puff of air, no higher than 50psi,
directed at the sensor housing. Let dry for at least
1 hour.
4. Replace the o-ring and lubricate with vacuum oil.
5. Check the oil catch filter for oil by blotting it on
a clean tissue or paper towel. If any oil remnants
appear, discard the oil catch lter and replace with
a new one.