Datasheet

6
Testing voltage : 3.2V peak (max.). The (+) socket is always positive relative to the (-) socket.
Input protection : The tester is protected against capacitor discharge (above 50V) by a 0.25A, 250V quick-acting
fuse.
Zeroing for capacitance : restricted to approx. ± 20pF
(2) Resistance
Range Resolution Accuracy
Max. open circuit
voltage
Overload
protection
200 0.1
±(0.75%rdg + 4dgt) 3.2V
500 Vd.c./a.c. (max)
2k 1
±(0.5%rdg + 1dgt)
20k 10
200k 100
2M 1k
±(0.75%rdg + 1dgt)
20M 10k
±(2%rdg + 1dgt)
0.5V