Datasheet

AC A / DC A
Volt Sense
The number of dashes displaying on the LCD
indicates the electric field intensity.
If no indication, voltage could still be present.
Using only red test probe to work as mains
voltage indicator.
V
S
E
L
E
C
T
H
O
L
D
R
A
N
G
E
V
HOLD RANGESELECT
Disconnect
Battery Replacement
Refer to the following figure to replace the
Battery :
Carry Case Assembling
.9.
.10. .8.
.11.
Overvoltage category :
IEC 61010-1 300V CAT.. , 600V CAT.II.
CAT. Application field
The circuits not connected to mains.
The circuits directly connected to
Low-voltage installation.
The building installation.
The source of the Low-voltage installation.
Operating altitude : 2000m (6562 ft)
Pollution degree : 2
EMC : EN 61326-1
Dimensions (WxHxD) : 56 x 12 x 112mm
Weight : 115g
Accessories : Battery (installed), carry case
and Instruction Card.
Function Range Resolution Accuracy
ACA
400.0µA
0.1µA
4.000mA
0.001mA
± (1.5%+ 5 D)
1-3 Electrical Specifications
Accuracy is ±(% reading + number of digits) at
23°C ± 5°C < 80%RH.
Frequency Response : 50 ~ 500Hz
AC Conversion Type : AC Coupled Average
Sensing, RMS Indication.
Input Impedance : 10M, <100pF.
Overload protection : 600V rms.
Input Impedance : 10M, <100pF.
Overload protection : 600V rms.
Function Range Resolution Accuracy
ohm
400.0 0.1
± (0.9%+ 5 D)
4.000K 0.001K
± (0.9%+ 2 D)
40.00K 0.01K
400.0K 0.1K
4.000M 0.001M
40.00M* 0.01M
± (1.5%+ 5 D)
Open Circuit Voltage : 0.4V
* There is a little rolling less than
± 2%
Overload protection : 600V rms.
.12.
CONTINUITY CHECK
Continuity Threshold : Approx. <50
Continuity Indicator : 2.7KHz Tone Buzzer.
Input Protection : 600V rms.
Function Range Resolution Accuracy
Cap
40.00nF
0.01nF
±(5%+ 0.4nF )
400.0nF
0.1nF
±(2.9%+ 5 D)
4.000µF 0.001µF
40.00µF 0.01µF
400.0µF 0.1µF
Measuring Time : <30sec. (400.0µF range),
<10sec. (40.0µF range), <3sec. (other range)
Overload protection : 600V rms.
Function Range Resolution Accuracy
Hz
40.00Hz 0.01Hz
±(0.3%+ 5 D)
400.0Hz 0.1Hz
4.000KHz 0.001KHz
40.00KHz 0.01KH
Sensitivity : > 10Vp-p (40KHz range)
> 1.5Vp-p (other range)
The signal must have positive and negative
waveform of a cycle.
Max.Sensitivity : 600V rms.
Overload protection : 600V rms.
Function Range Resolution Accuracy
%
0.1 - 99.9% 0.1%
± (0.5%+10 D) *
* : 30% Duty 70%,
Square Wave (5 Hz ~ 1KHz)
Sensitivity : 1.5Vp-p
Overload protection : 600V rms.
Limited Warranty
This Meter is warranted to the original purchaser
against defects in material and workmanship for
two years from the date of purchase. During this
warranty period, manufacturer will, at its option,
replace or repair the defective unit, subject to
verification of the defect or malfunction. This
warranty does not cover fuses, disposable batter-
ies, or damage from abuse, neglect, accident,
unauthorized repair, alteration, contamination, or
abnormal conditions of operation or handling.
Any implied warranties arising out of the sale of
this product, including but not limited to implied
warranties of merchantability and fitness for a
particular purpose, are limited to the above. The
manufacturer shall not be liable for loss of use of
the instrument or other incidental or consequential
damages, expenses, or economic loss, or for any
claim or claims for such damage, expense or
economic loss. Some states or countries laws
vary, so the above limitations or exclusions may
not apply to you.
.13. .14.
.7.
SELECTHOLD RANGE
Specifications
1-1 General Specifications
LCD display digits :
4000 counts digit large scale LCD readout.
Measuring rate : 3 times / sec.
Polarity Indication :
Automatic, positive implied,
Overrange display : “OL” or “-OL”
Unit symbol indication.
Automatic power off time :
Approximately 20 minutes after power on.
Low battery indicator : < is displayed.
Power requirement : 1.5V x 2 batteries.
Battery life : 50 hours (GPA76P)
1-2 Environmental Conditions
Indoor Use.
Calibration : One year calibration cycle.
Operating temperature :
0°C ~ 30°C (80% RH)
30°C ~ 40°C (75% RH)
40°C ~ 50°C (45%RH)
Storage temperature :
-20 to +60°C, 0 to 80% RH (batteries not fitted).
Temperature coefficient : Add 0.2 x (Specified
accuracy) / °C, < 18°C, > 28°C .
Shock vibration : Sinusoidal vibration per
MIL-T-28800E (5 ~ 55 Hz, 3g maximum).
Drop Protection : 4 feet drop to hardwood on
concrete floor.
Function Range Resolution Accuracy
ACV
400.0mV 0.1mV
± (1.5%+ 5 D)
4.000V 0.001V
40.00V 0.01V
400.0V 0.1V
600V 1V
± (0.9%+ 5 D)
Function Range Resolution Accuracy
DCV
400.0mV 0.1mV
± (0.7%+ 5 D)
4.000V 0.001V
40.00V 0.01V
400.0V 0.1V
600V 1V
± (0.7%+ 5 D)
± (0.6%+ 2 D)
Frequency Response : 50 ~ 500Hz
AC Conversion Type : AC Coupled Average
Sensing, RMS Indication.
Input Impedance : Approx. 3K
Overload protection : 600V rms.
Function Range Resolution Accuracy
DCA
400.0µA 0.1µA
4.000mA 0.001mA
± (0.9%+ 5 D)
Input Impedance : Approx. 3K
Overload protection : 600V rms.
DIODE TEST
Test Current : Typical. 350µA
Open Circuit Voltage : Max. 1.8V DC
Input Protection : 600V rms.