AC/DC CURRENT TRANSDUCER INSTRUCTION MANUAL APPA 39T
INTRODUCTION 1-1 Unpacking and Inspection Upon removing your new Clamp Meter from its packing, you should have the following items: 1. Current Transducer with coiled cable output plugs. 2. Carrying Case. 3. Instruction manual. 1-2 Meter Safety Terms as Marked on Equipment ATTENTION — Refer to manual DOUBLE INSULATION — Protection Class II. DANGER — Risk of electric shock.
Symbols in this Manual = This symbol indicates where cautionary or other information is found in the manual. Battery 1-3 Front Panel Refer to Figure 1 and the following numbered steps to familiarize yourself with the meter’s front panel controls and Connectors. 1. Red LED — Low battery display. 2. Green LED — Power on display. 3. Coiled Cable Output Leads — Voltage output. 4. Function Switch — Function select switch used to select measuring function. 5.
Figure 1 3
SPECIFICATIONS 2-1 General Specifications Position Error : +/- 1% of Reading. Type Of Sensing : Hall effect sensing for AC and DC. Low Battery Indicator : Red LED indicator. Power On Indicator : Green LED indicator , flashing. Temperature Coefficient : 0.2 x (Spec.Acc’y) / °C , < 18°C or > 28°C . Power Requirement Battery Life Max / Conductor Size : Single 9V battery (NEDA 1604A or IEC 6LF22) (Use supplied) : 66 hours typical (Alkaline). : 51mm diameter or 24 x 60mm busbar.
2-2 Environmental Conditions : Indoor use. Maximum Altitude Installation Category Pollution Degree : 2000 Meter : IEC 1010 1000V CAT. III. :2 Operating Ambient Storage Temperature : 0ºC ~ 50ºC , <75%R.H. : -20ºC ~ 60ºC, < 80% R.H.
2-3 Electrical Specifications All at 23ºC ± 5ºC, less than 75% R.H. Ranges : 0 ~ 100A/1000A AC 0 ~ 100A/1000A DC Accuracy (1) DC Current : Ranges : 100A, 1000A Output Voltage : 10mV / A for 100A Range 1mV / A for 1000A Range Accuracy : 100A Range : (0 - 100)A ±(2.9% + 2A) 1000A Range : (100 - 400)A ± (1.9% + 7A) 1000A Range : (400 - 1000)A ± (2.
(2) AC Current : Ranges : 100A, 1000A Output Voltage : 10mV / A for 100A Range 1mV / A for 1000A Range Accuracy : 100A Range : (0 - 100)A ±(1.9% + 2A) 40-400Hz 1000A Range : (100 - 400)A ± (1.9% + 7A) 40-400Hz 1000A Range : (400 - 1000)A ± (2.9% + 5A) 40-400Hz Load Impendance : 100KΩ min.AC/DC (3) Auto Power Off : Approximately 30 minutes after function changes.
OPERATION This instrument has been designed and tested in accordance with IEC Publication 1010, Safety Requirements for Electronic Measuring Apparatus and has been supplied in a safe condition. This instruction manual contains some information and warnings which have to be followed by the user to ensure safe operation and to retain the instrument in safe condition. 3-1 Precautions and Preparations for measurement 1. Do not apply the voltage to the output plugs. 2.
THIS INSTRUMENT MUST NOT BE USED ON UNINSULATED CONDUCTORS AT A VOLTAGE GREATER THAN 1000V ac/dc. 3-2 AC Current Measurement 1. Select the highest anticipated input (1000A or 100A) by the function switch. 2. Insert the “Output plugs (red and black)” to the input terminals of a Multimeter. Set the Multimeter to “AC 200mV or 2V” range.
3-3 DC Current Measurement 1. Select the highest anticipated input (1000A or 100A) by the function switch. 2. Insert the “Output plugs (red and black) “ to the input terminals of a Multimeter. Set the Multimeter to “DC 200mV or 2V” range. 3. Be sure to zero adjust first. With jaws closed and away from conductor, continue to press DCA AUTO ZERO Button until instrument reads all zeroes or very close to it. 4.
CURRENT Figure 2 11
MAINTENANCE TO AVOID ELECTRICAL SHOCK REMOVE TEST LEADS BEFORE OPENING THE COVER. General Maintenance 1. Repairs or servicing not converted in this manual should only be performed by qualified personal. 2. Periodically wipe the case with a dry cloth and detergent do not use abrasives or solvents. Battery Installation Replacement The meter is powered by a single 9V battery. Refer to Figure 3 and use the following procedure to replace the battery. 1. Disconnect the output plus and turn the meter off. 2.
9V Battery 9V Battery Figure 3 13
APPA TECHNOLOGY CORP. 9F.119-1 Pao-Zong Rd., Shin-Tien, Taipai, 23115, Taiwan, R.O.C. P.O.Box. 12-24 Shin-Tien, Taiwan. Tel : 886-2-9178820 Fax : 886-2-9170848 E-MAIL:info @appatech.com http://www.appatech.com Printed In Taiwan Copright 2001, APPA Tech., Corp. All rights reserved.