Specifications

Vessel
4
Caution! The positive pressure relief and seal-off
device protects the vacuum room from
overpressure. Re-evacuation may only be carried
out by
manufacturers skilled staff
The protective cover (2) intercepts the valve insert
(1), when there is overpressure existing in the
vacuum room.
Do not remove the protective cover (2).
Protect the valve from heat as well as cooling, as
brittleness results in the loss of the operating
vacuum.
Structure of the Siphon
Item
Description
1 Waste gas/overflow line
2 Basic body with small flange connection
DN 50
3 Waste gas/overflow valve G 1/2
4 Filling/withdrawal valve G 3/8
5 Connection screwing
(double nipple ring R 3/8 - 3/4-16 UNF)
for flexible transfer hose
6 Filling/withdrawal line
The siphon serves for the filling and withdrawal of liquid
nitrogen.
Special designs (e.g. single / triple withdrawal) are available
on request.
Ball Valve Positions:
A - closed
B - open
2.7 Combined Positive Pressure
Relief and Seal-off Device
2.8 Siphon with Small Flange
Connection
Type EK