Instruction manual

30
(Continued on page 31)
31
CALIBRATION continued from page 29
CALIBRATION continued from page 30
AUTO CAL automatically calibrates the instrument to all gases
detected by the sensors installed. From the AUTO CAL menu,
press the TICK/MENU BUTTON (B) . Attach the prepared gas
as listed on the display. A countdown timer shows the time re-
maining to connect . Failure to connect will result in a FAIL
message and an alarm sound (press any button to continue).
During the test, the gas type, the target concentration and digi-
tal output of the sensor (not concentration) will be displayed.
Following each gas tested PASS or FAILED will be displayed. It
is possible to manually stop the test once it is started by press-
ing the POWER/MUTE BUTTON (A).
At the end of the test BUMP TEST (PASS or FAILED) will be
indicated. Pressing the POWER/MUTE BUTTON (A) will save
the results and return to the USER MENU.
Pressing the SAVE/ZERO BUTTON (C) will scroll through the
individual sensors that are configured in your G2 for calibration.
Pressing the TICK/MENU BUTTON (B) selects the option. At-
tach the proper gas and wait for the instrument to provide a
PASS or FAILED indication.
If the calibration has failed press the POWER/MUTE BUTTON
(A) to attempt again or scroll to the next gas. Upon successful
calibration the calibration due date reminder is automatically
updated.
From this menu pressing the SAVE/ZERO BUTTON (C) again
will provide access to the SMART-CAL automatic calibration
system. Pressing the TICK/MENU BUTTON (B) from SMART-
CAL menu begins the communication process. Consult the
SMART-CAL system instructions for connection details.
NOTE: A calibration failure is indicated on the display by
FAILED. Re-calibration should be attempted. Any instru-
ment that does not accept calibration should be taken out
of service. Please contact SENSIT TECHNOLOGIES for any
needed repairs.