Instruction manual

MENU continued from page 17
(Continued on page 19)
18
19
MENU continued from page 18
(Continued on page 20)
SET TIME
From the working display access the menu by pressing and
holding the TICK/MENU BUTTON (B) until the top line reads
USER MENU SHOW TIME. Press & release the SAVE/ZERO
BUTTON (C) to scroll to USER MENU SET CLOCK.
Press the TICK/MENU BUTTON (B) to select this feature. The
day of the month will flash. Use the TICK/MENU BUTTON (B) to
change the date. Using the SAVE/ZERO BUTTON (C) will scroll
to the next item to change. Press the POWER/MUTE BUTTON
(A) to save the changes and return to the USER MENU.
PRINT
From the working display access the menu by pressing & hold-
ing the TICK/MENU BUTTON (B) until the top line of the display
reads USER MENU SHOW TIME. Press the SAVE/ZERO BUT-
TON (C) to scroll to USER MENU PRINT MENU. Press the
TICK/MENU BUTTON (B) to select this feature. PRINT MENU
SESS LOG will be displayed and the pump will turn off. Press
the SAVE/ZERO BUTTON (C) to scroll to the type of report to
print.
Prepare and align the printer. Use the TICK/MENU BUTTON (B)
to begin the printing. Do not block or interfere with the line of
sight between the printer and the instrument during the printing
time. Press POWER/MUTE BUTTON (A) to return to the USER
MENU. The pump will also restart.
BUMP TEST
Prior to starting this test prepare the necessary gases per the
sensor configuration.
From the working display, pressing and holding the TICK/MENU
BUTTON (B) until the display reads USER MENU SHOW TIME.
Press the SAVE/ZERO BUTTON (C) to scroll to USER MENU
BUMP TEST. Press the TICK/MENU BUTTON (B) to select this
feature.
Attach the prepared gas. During the test the gas type, the target
concentration, the time remaining in seconds and actual con-
centration will be displayed. After each gas is tested PASS or
FAILED will be displayed. It is not possible to manually stop the
test once it is started. At the end of the test BUMP TEST
(PASSED or FAILED) will be indicated. Press the POWER/MUTE
BUTTON (A) will save the results and return to the USER MENU.
SMART-CAL
From the working display, press & hold the POWER/MUTE BUT-
TON (A) for 2-3 seconds. The display will read SMART CAL
communicating.
Place the instrument into the cradle on the left side of the Smart-
Cal Calibration Station. Attach the tubing from the station to the
instrument. Press & release the TICK/MENU BUTTON (B), the
display will show SMART CAL communicating.
Press & release the CALIBRATE button on the Smart -Cal and
calibration will begin automatically. If successful, CALIBRATION
PASSED will show on display. If unsuccessful, CALIBRATION