User`s manual
Operation 2011-09-30
iv
WARNI N G
• This unit must not be connected to or used in combination with any other apparatus or system. It
must not be used as an integral component of any other apparatus or system.
J. Morita Mfg.
Corp. will not be responsible for accidents, equipment damage, bodily injury or any
other trouble which results from ignoring this prohibition.
• Accurate canal measurement is not always possible depending on the shape and condition of the
tooth as well as a decline in the equipment’s performance.
• Do not use damaged file holders; an accurate measurement can not be made with a damaged file
holder.
• When a continuous tone is heard while the main power switch is on and without any operation,
some electrical part may be malfunctioning. Do not use the unit and send the unit to J. Morita
regional office for repairing.
• A rubber dam should be used when performing endodontic treatment.
• Some care must be taken concerning electromagnetic compatibility (EMC) when using the Root ZX
mini. Refer to the user’s manual and other attached documents for EMC information regarding
installation and operation.
• Both portable and movable radio frequency transmitters may have some effect on the Root ZX
mini.
• Using replacement parts or accessories not supplied by the original manufacturer or vender could
adversely affect the EMC performance of the Root ZX mini.
• As far as possible, do not use the Root ZX mini near or simultaneously with other devices. If this
cannot be avoided, observe carefully and make sure both the Root ZX mini and the other device
operate normally.
:This indicates when not to use the equipment.
• Do not use this unit in conjunction with an electric scalpel or on patients who have a pacemaker.
• Blocked canals cannot be accurately measured.
• This unit must not be connected to or used in combination with any other apparatus or system.
It must not be used as an integral component of any other apparatus or system.
J. Morita Mfg.
Corp. will not be responsible for accidents, equipment damage, bodily injury or any
other trouble which results from ignoring this prohibition.
• Illumination devices such as fluorescent lights and the Film viewer which use an inverter can cause
the Root ZX mini to operate erratically. Do not use the Root ZX mini near devices such as these.
• Electromagnetic wave interference could cause this device to operate in an abnormal, random
and possibly dangerous manner. Cellular phone, transceivers, remote controls and all other
devices which transmit electromagnetic waves located inside the building should be turned off.