User`s manual

Operation 2011-09-30
17
Storage
Transport and Storage Conditions:
Temperature: -10° C (14° F) to 70° C (158° F), Realtive Humidity: 8 ~ 80 % (without
condensation), Atmospheric Pressure : 700 ~ 1,060 hPa
Do not expose to x-rays or direct sunlight frequently or for long times.
If the unit has not been used for a long time, make sure it works properly before using.
Always remove the batteries prior to storing or shipping the unit.
The J. Morita Mfg. Corp. will supply replacement parts and be able to repair the product for a period
of 10 years after the manufacture of the product has been discontinued.
7.
Maintenance and Inspection
Maintenance and inspection are generally consider to be the duty and obligation of the user, but if, for some
reason, the user is unable to carry out these duties, he may rely on a qualified medical device serviceman.
Contact your local dealer or the J. Morita Corp. for details.
Replace the parts listed in the Parts Lists as necessary depending on degree of wear and length of use.
This apparatus should be inspected every 6 months in accordance with the following maintenance and inspection
items.
Maintenance and Inspection Items
1. Check that the Power switch turns the unit on and off properly.
2. Insert the Tester and check that the indicator is within ±3 lines of 1 on the meter.
3. Check that the Set switch changes the memory from 01 to 02 to 03.
4. Check that the probe cord can be properly plugged into its jack.
5. Check that the file holder's plug can be connected properly to the probe cord and that the file
holder can be clipped onto a file.
Check the contrary electrode can be plugged into its probe cord connector.
6. Touch the contrary electrode with the file holder and make sure all the bars on the meter light up.
7. This unit should be inspected after prolonged unusual period.
Parts Lists
Probe Cord File Holder Contrary Electrode
Code No. 8449627
Code No. 7503670 Code No. 7503680
Tester Long File Holder
Code No. 8449430
Code No. 7503673