Sterilization and Replacement Parts
Sterilization
a. Autoclavable Components [File Holder, Long File Holder(Option) and Contrary Electrode]
Recommended temperature and time: 135°C (275°F), 4 minutes minimum with a sterilization pouch.
Minimum drying time after sterilization: 10 minutes.
or Recommended temperature and time: 121°C (249.8°F), 35 minutes minimum with a sterilization
pouch. Minimum drying time after sterilization: 30 minutes.
Autoclaving and drying temperatures must never exceed 135° C/ 275° F.
The file holder, long file holder and contrary electrode must be thoroughly washed and cleaned before
autoclaving.
Any chemicals or foreign debris left on instruments could cause them to malfunction or could cause
discoloration.
WARNI N G
Autoclave file holder, long file holder and contrary electrode after each patient.
NOTE
Do not sterilize in any way other than autoclave.
It is highly recommended that instruments be autoclaved in a sterilization pouch (wrapped) or similar
device
.
Do not autoclave probe cord.
Follow the manufacturer’s recommendations to disinfect files.
b. Sterilize Surface of Main Unit and Probe Cord with Ethanol
* Wipe with a piece of gauze dampened with Ethanol for Disinfection (Ethanol 70~80 vol%).
Wring the gauze out to make sure it is not too wet.
Do not use an excessive amount of Ethanol for Disinfection (Ethanol 70~80 vol%); it could seep inside and
cause a malfunction. Be especially careful when cleaning around the port for data output and the jack for the
probe cord.
Use only Ethanol for Disinfection (Ethanol 70~80 vol%) for disinfection. Any other type of solution could cause
cracking, cloudiness, discoloration or some similar kind of damage.
NOTE
On rare occasions, static electricity produced by wiping the liquid crystal display with a dry cloth may
have some influence on the appearance of the display.
Avoid spilling chemical solutions used for treatment on the Root ZX mini. These chemicals could cause
damage, deform or discolor the Root ZX mini. Be especially careful to avoid spilling formalin cresol
(FC) and sodium hypochlorite as they are quite strong. Wipe up any chemical spills immediately.
(Some chemicals may leave discoloration and spots even if they are immediately wiped up.)
Replacement Parts
* Replace the parts as necessary depending on degree of wear and length of use.
* Order parts from your local dealer or the J.Morita Corp.