User`s manual
Operation 2011-09-30
10
Rubber Stoppe
r
Set Switch
File Holder
0.5 Reading
Flash Bar
4. Press the Set Switch to select Memory 01, 02 or 03.
5. Insert the file up to the Flash Bar (this point can also be
recognized by the change in the beeping). Position the
rubber stopper on the tooth surface as a reference point to
determine the root canal’s working length. Use the 0.5
reading on the meter to estimate the canal’s length.
6. Determine the working length.
■ If the file tip is at the 0.5 meter reading, subtract from
0.5 to 1.0 mm to determine the working length.
* The working length will differ somewhat depending on
each individual tooth. This discrepancy must be
judged by the dentist as he works on the tooth.
NOTE
Make sure to take an x-ray to check the results.
■
When using the long file holder instead of the file holder
NOTE
Make sure the long file holder does not prick or pierce the
patient’s oral mucosa .
Long File Holder
(Option)