WARNI N G
Check the Root ZX mini’s operation before each patient. If
the indicators in the display do not all appear normally, the
instrument may not be able to make an accurate
measurement. In this case, stop using the instrument and
have it repaired.
Checking the Function with the Tester
Check Root ZX mini’s performance with the tester once a
week.
1. Press the Power switch to turn the unit on.
2. Insert the tester into the probe cord jack.
Check that the meter indicates within ±3 bars away from
(above or below) 1.
* The meter may jump when the tester is inserted. If it
does, wait for about one second until the meter stabilizes
and then check the reading.
* If the reading is 4 or more bars away from 1, the unit will
not make an accurate measurement. In this case, contact
your local dealer or J. Morita regional office.