Specifications
6 
English   
4.  Cleaning 
4.1  Compressor Outer Case Cleaning 
WARNING - 
Electric shock  hazard. Do  not  remove  outer  case of  this  unit.  All  disassembly  and 
maintenance of this unit must be done by a qualified service technician. Refer servicing 
to qualified service personnel. 
WARNING - 
This unit does not require oil. Do not attempt to lubricate any internal parts. 
WARNING -
 Unplug unit before cleaning. Do not submerge in water for cleaning.
1.  Wipe the main unit with a damp cloth every few days to keep it dust-free. 
2.  Do not use any powdered type cleaners or soap. Do not submerge the unit into water. 
4.2  Nebulizer Cup Cleaning 
All parts of the disposable nebulizer, except tubing, should be thoroughly cleaned after each use. If your doctor 
specifies a specific cleaning procedure, follow that procedure. 
WARNING - To prevent possible risk of infection from contaminated  medication, cleaning of the 
nebulizer is recommended after each treatment. 
1.  After each use, disassemble the  nebulizer cup, tee and mouthpiece completely, wash all parts in warm 
soapy water and rinse thoroughly. Allow to air dry. Store in a clean dry location. 
2.  Once daily, after washing as above, soak nebulizer components in a fresh solution consisting of one part 
warm water and one part white vinegar for 30 minutes followed by a warm water rinse. 
NOTE - 
The disposable nebulizer kit is for single patient use only. 
NOTE -
 Please follow national requirements to dispose the unit properly. 
5.  Storage 
Keep the unit and accessories dry. Avoid direct sunshine. 