Specifications
5
English
3. Operation
NOTE -
Before initial operation, the nebulizer cup assembly should be cleaned following instructions
described in the “Cleaning” section.
WARNING -
Before connecting the power cord, make sure the I/O (ON/OFF) switch is in the O (OFF)
position.
The plug is also served to disconnect the device.
3.1 Daily Use Operation
CAUTION -
The Mini-Plus Compressor is designed for intermittent use only. Do not operate it
continuously for more than 30 minutes for a single use without turning it off and
following a cooling period for least 30 minutes.
1. Before each use inspect the Mini-Plus Compressor and nebulizer cup assembly for damage or wear,
replace as needed.
2. Place the Mini-Plus Compressor on a table or other flat stable surface. Be sure you can easily reach the
controls when seated. Do not use this device on the floor.
3. With the power switch in the O (OFF) position, plug the power cord into an appropriate electrical wall
outlet.
4. Connect one end of the tubing to the compressor air-outlet connector.
5. Assemble the nebulizer cup and add indicated medication to the nebulizer’s cup before use.
6. Attach the other end of the tubing into the air-inlet connector founded at the bottom of the nebulizer cup.
7. Turn on the Mini-Plus Compressor by pressing the power switch to the I (ON) position and begin
treatment.
8. If treatment needs to be interrupted, simply press power switch to O (OFF) position.
9. When the treatment is complete, turn off the compressor by pressing the power switch to O (OFF) position
and unplug the unit from the electrical outlet.
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or
with oxygen or nitrous oxide. This device has no AP or APG protection.
3.2 Safety Overload
NOTE - Do not exceed 30 minutes of continuous operation.
1. The motor of this device has a built-in thermal overload protector. Should the motor overheat, the protector
will automatically shut off the motor. Should this occur, turn the I/O (ON/OFF) switch to the O (OFF)
position and allow the motor to cool down for approximately 30 minutes before turn it on again.
2. If the overload protector shuts off the motor frequently, you may have an unstable voltage situation.
3. If the unit shuts down and cannot restart, it may need to be replaced. Call your provider immediately.