Appendix A: EMC Information
Guidance and Manufacturer’s Declaration- Electromagnetic Emissions:
This device is intended for use in the electromagnetic environment specified below. The user of this device should make
sure it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment-Guidance
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations /
Flicker emissions
IEC61000-3-3
Complies
The device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network.
Guidance and Manufacturer’s Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The user of this device should make
sure it is used in such an environment.
Immunity Test
IEC60601 test
level
Compliance Electromagnetic Environment-Guidance
Electrostatic Discharge (ESD)
IEC61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30 %.
Electrical fast transient/ burst
IEC61000-4-4
±2kV for power supply
line
±1kV for input/out line
±2kV for power supply
line
±1kV for input/out line
Mains power quality should be that of atypical
commercial or hospital environment
Surge
IEC61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to
line(s)
Mains power quality should be that of atypical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC61000-4-11
<5 % U
T
(>95 % dip in
U
T
)for 0,5 cycle
40 % U
T
(60 % dip in
U
T
)for 5 cycles
70 % U
T
(30 % dip in
U
T
)for 25 cycles
<5 % U
T
(>95 % dip i
n
U
T
)for 5 sec
<5 % U
T
(>95 % dip in
U
T
) for 0,5 cycle
40 % U
T
(60 % dip in
U
T
) for 5 cycles
70 % U
T
(30 % dip in
U
T
) for 25 cycles
<5 % U
T
(>95 % dip in
U
T
) for 5 sec
Mains power quality should be that of atypical
commercial or hospital environment. If the user of this
device requires continued operation during power mains
interruptions, it is recommended that the device be
powered from an uninterruptible power supply or a
battery.
Power frequency
(50/60Hz) magnetic field
IEC61000-4-8
3 A/m 3 A/m
Power frequency magnetic fields should be at levels
characteristic of atypical location in a typical commercial
or hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to the application of the test level