Specifications
9
English
8. Specifications
Electrical Rating
(Note: Refer to the rating label on
the product)
120VAC,60Hz,1.2A
Maximum Compressor Pressure
35 psi
Operation Flow Rate 5.5lpm
Nebulization Rate 0.2~ 0.3 ml/min*
Particle size 0.5 ~ 5 microns ( m)*
Classification
Class II. BF equipment.
IPXO. No AP/APG protection.
Applied part Mouthpiece or Nasal mask
Dimensions (W x D x H)
14.0 18.8 10.0 cm / 5.5” 7.4” 3.9”
Weight
1.6 kg / 3.5 lb
Fuse T5A 250VAC for 120V system
Temperature
Operation: 10º C to 40º C / 50º F to 104º F
Storage: -15º C to 50º C / 5º F to 122º F
Transport: -15º C to 70º C / 5º F to 158º F
Environment
Humidity
Operation: 10% to 90% RH non-condensing
Storage: 10% to 90% RH non-condensing
Transport: 10% to 90% RH non-condensing
Note:
1.
*
Values are determined by supplier’s nebulizer.
2. Consult distributor(s) or EU representative(s) for additional technical documents.
3. The Nebulizer is fitted with a fuse as specified in the Specification column, it is designed to be maintained by
the manufacturer or its agents. Users are not allowed to open the cover to replace a blown fuse.