How-To Guide
GlucoSure Link Blood Glucose Monitoring System
54
Cleaning and disinfection frequency: The GlucoSure Link Blood Glucose meter and
lancing device should be cleaned when it appears dirty or contaminated with blood.
It should be disinfected every day according to the validated cycle of 1825 times.
The validated cycle of 365x5=1825 times is equivalent to one cleaning plus one
disinfection per day over 5 years.
Cleaning and disinfecting your meter and lancing device
Clean and disinfect your meter and lancing device at least 1 time per week using
DISPATCH hospital cleaner disinfectant towels with bleach, with EPA registration
number 56392-8. Other wipes with EPA Reg. No. 56392-8 may also be used.
1. Wash hands with soap and water.
2. Inspect for blood, debris, dust, or lint anywhere on the meter.
3. Clean or disinfect meter or lancing device.
4. To clean the GlucoSure Link Blood Glucose meter or lancing device, blood/body
uids must be thoroughly cleaned from surface and objects before disinfecting.
Dispense the towel (DISPATCH) from its package. Wipe surface with towel
(DISPATCH) until completely wet. You should wipe all external areas of the meter
including both front and back surfaces. Allow to remain wet for one minute at room
temperature (68-77°F). Discard used towel in a sealed container where it will not
be touched by others.