How-To Guide

GlucoSure Link Blood Glucose Monitoring System
30
Testing Your Blood Glucose
You will need:
Your GlucoSure Link Blood Glucose Meter
A new GlucoSure Link Blood Glucose Test Strip
Lancing device with a clean, unused lancet (or safety lancets)
Clear cap for Alternate Site Testing (AST) on palm and forearm
Before you begin, make sure:
To set up your meter properly and run a control test. See Setting Up Your New
System and Control Solution Testing for details.
Wash your hands and the testing site thoroughly with soap and warm water, and
dry well.
You are testing in an area between the temperature ranges of 50°F~104°F. Your
meter will not test outside of this range and will display thermometer icon. Move the
meter into an area that is between 50°F~104°F, and let it sit for 10 to 15 minutes
before testing again.
If the meter is being operated by a second person who is providing testing
assistance to you, the meter and lancing device should be cleaned and disinfected
appropriately when testing is conducted by the second person.