How-To Guide
GlucoSure Link Blood Glucose Monitoring System
27
Step 4.
Squeeze a drop of control solution onto a clean, dry, nonabsorbent
surface.
Do not apply control solution to the test strip directly from the bottle.
Replace the bottle cap on the control solution bottle immediately
after use.
Step 5.
Gently touch the drop of the control solution to the tip of the test strip.
The meter will beep indicating it has received the control solutio
Step 6.
The screen will start to count down.
After 6 seconds, the control test result will appear on the meter’s
display screen.
Step 7.
Compare the reading on the screen to the Level 1 range printed
on the test strip bottle or its package. If the reading falls outside
the Level Low or Level 1 range printed on the test strip bottle or its
package, see Control Solution Trouble Shooting, on page 29.
L1:
L2:
Level Low :
Level 1:
Level 2:
43 - 75
96 - 135
179 - 242