• Keep away from direct sunlight and heat. Store the test strip bottle in a dry, cool
place.
• Record the "date opened" on the bottle label. Discard the bottle and any remaining
test strip after 3 months from date of opening.
• Test under the temperature range of 50°F to 104°F, relative humidity 20 to 85%RH.
Your meter will not work outside these ranges.
• Must insert test strip into the test strip port. .
• Must see a ashing blood drop icon if the test strip has been inserted to the test
strip port.
Step 2.
Wait until the ashing blood drop and arrow icons appear on the
meter display screen. Press ▼ to enter Low or L1 control solution
testing.
Step 3.
ctl icon will appear next to the test strip icon and L1 will appear on the
meter display screen, please apply control solution onto the test strip.
If you decide not to do a control solution test, Press ▲ until ctl disappeared from the
meter display.
mg /dL