How-To Guide

GlucoSure Link Blood Glucose Monitoring System
25
unused control solution 3 months after opening.
DO NOT FREEZE. Store the control solutions at 59°F to 86°F.
Do not drink the control solution. It is not intended for human consumption.
Avoid contact of the solution with the skin and the eyes as this could cause
inammation.
Discard used control solution bottles according to your local regulation.
You will need:
Contrex™ Plus III Control Solution (Low, L1 and L2)
Your GlucoSure Link Blood Glucose Meter
A new GlucoSure Link Blood Glucose Test Strip
Step 1.
Take out a GlucoSure Link Blood Glucose Test Strip from the test
strip bottle and close the bottle immediately. Insert the test strip to
turn on the meter.
IMPORTANT:
Check the expiration date on the test strip bottle and its package. Do not use
expired test strips.
Use each test strip immediately after removing it from the bottle.
Do not use wet or damaged test strips.