Instruction Manual
Factory default is Aut.
b) P2-Select temperature unit (°C-°F)
In P1 mode, press to enter P2, press , °C or °F flickering, press to
choose °C-°F, press to confirm parameter selection. Long press to return to
measurement mode.
c) Restore to factory default:
Select Yes to restore the calibration to the theoretical values and parameter setting to
original values. When meter’s calibration or measurement performs abnormally, this
function can be adopted so the meter goes back to factory default setting and then users
can conduct calibration or take measurements again.
8. Technical Specifications
Conductivity
Range
0~199.9 µS/cm, 20~1999µS/cm,
2.0~20.00mS/cm
Resolution
0.1/1 µS/cm, 0.01 mS/cm
Accuracy
±1% F.S±1 digit
Calibration points
1-2 points auto calibration
Automatic Temp.
Compensation
0-50˚C
Temp.
Range
0-50˚C
Resolution
0.1˚C
Accuracy
±0.5˚C
9. Other Functions & Parameters
Indication of
calibration points
Auto Power-off
Power-off in 8 minutes if
no operation
Indication of stable
measurements
Waterproof level
IP67, floats on water
Self-Diagnosis
information
Er1, Er2
Power Supply
AAA batteries*4
Low battery reminder
Flashes to
remind to replace
batteries
Battery Life
Up to 1000 hours of
continuous operation
Dimensions/Weight
Instrument: 40*31*178mm/107g;
Carrying case: 190*165*140mm/438g