User Manual

301Pt-C ORP Electrode
User Manual
Brief Introduction
Apera Instruments 301Pt-C ORP Electrode is featured with a proprietary platinum ring ORP sensor for
fast and stable readings. The gel inner reference solution does not need to be refilled. This electrode is
designed for testing in general water solutions both in-lab or in-field.
Technical Specifications
Range
-1999 to 1999 mV
Sensor
Φ6×2.5 platinum ring
Housing
Polycarbonate
Junction
Single ceramic
Reference Electrode
Ag/AgCl
Connector
BNC
Cable length
3 Ft
Dimension
ø12*160 mm
Standard value
222 mV +/- 15 mV (at 25°C)
Operating temperature
32 to 176°F (0 - 80°C)
How to Install the Electrode
1. Find the BNC socket (where it shows pH ORP) on the pH meter; Open the rubber cap; Insert the
blue BNC connector of the electrode to the BNC socket while twisting clockwise until it’s locked.
2. There is no calibration needed for ORP electrodes. If you want to double-check the condition of
the electrode, simply test the electrode in the standard ORP solution such as 222 mV. If the
reading is within the error range (+/- 15 mV), then it is good to go.
How to Use the Electrode
1. There is a proper amount of storage solution in the cap on top of the electrode. The ORP sensor
tip is soaked in it to keep its sensitivity.
2. Before measuring, loosen the bottle cap, then pull out the electrode while twisting
counterclockwise. Place the storage bottle at a safe position.
3. Rinse the electrode with purified water and shake off excess water or blot-dry it with clean tissue
or filter paper. Never rub the sensor.
4. Insert the electrode into your sample solution and stir it for a few seconds in the solution to
remove potential air bubbles, which may cause unstable readings. Then wait for the stable
reading and take the measurement.