- 8 -
4.7 Parameter Setting
Table -9
Symbol
Parameter setting content
Parameter
Factory default setting
P1
Select electrode constant
0.1, 1.0, 10
1.0
P2
Select reference temperature
15 - 30
25
P3
Adjust temperature compensation
coefficient
0.00 - 9.99%
2.00%
P4
Adjust TDS Factor
0.40 1.00
0.71
P5
Select temperature unit
-
P6
Clear stored data
No Yes
No
P7
Restore to factory default
NoYes
No
4.7.1 How to do the parameter setting:
In measurement mode, long press to enter parameter setting P1. Then press or to
change menu from P1 to P2…P7. For details, please see Table-10.
Table-10 pH Parameter Setting
P1 Select electrode’s constant1.010.00.1
1. Long press to enter P1
2. Press , “1.0” flashes, press to choose 1.0→10.0→0.1; press
to confirm
3. Press to enter P2, or press to return to measurement mode.
P2Select reference temperature (25.018.020.0 )
1. Press , “25.0 flashes, press or to adjust temp. from 15 to
30; press to confirm.
2. Press to enter P3, or press to return to measurement mode.
P3Adjust temp. compensation coefficient (0.00-9.99%)
1. Press , “2.00” flashes, press or to adjust from 0.00-9.99; press
to confirm.
2. Press to enter P4, or press to return to measurement mode.