User Guide

6
6. Parameter Setting
6.1 Setting Schedule
6.2 Parameter Setting
When turned off, long press to enter setup short press to switch
P1-P2-P3 Short press , parameter flashingshort press to choose,
short press to confirm parameter selectionLong press to go back to
measurement mode.
6.3 Parameter Setting Instruction
Select standard pH buffer solution (P1): There are two options of standard buffer
solutions: USA series and NIST series as the following chart:
Icons
pH standard buffer solution series
USA series
NIST series
Three-Point
calibration
4.00 pH
4.01 pH
7.00 pH
6.86 pH
10.01 pH
9.18 pH
Prompt
Mark
Parameter Setting Items
Code
Factory
Default
P1
Select pH buffer
USA NIST
USA
P2
Select Temperature Unit
˚F ˚C
/
P3
Back to Factory Default
No Yes
No