User Manual

10
Conductivity
Range
0 to 200.0 μS, 200 to 2000 μS,
2 to 20.00 mS/cm
Resolution
0.1/1 μS, 0.01 mS/cm
Accuracy
±1% F.S
Calibration Points
1 to 3 points
TDS
Range
0.1 ppm to 10.00 ppt
TDS Factor
0.40 to 1.00
Salinity
Range
0 to 10.00 ppt
Temperature
Range
0 to 50˚C (32-122˚F)
Resolution
0.1˚C
Accuracy
±0.5˚C
12
Probe Replacement
To replace a probe: 1) Take off the probe cap; 2) Screw off the probe ring 3)
Unplug the probe; 4) Plug in the new replacement probe (pay attention to the
probe’s position); 5) Screw on the probe ring tightly. Soak the probe in 3M KCL
for 5-15 minutes. Then perform calibration before testing.
Probes compatible with PC60 Tester:
AI1211 PC60-E (Regular pH/conductivity probe), AI3719-E PC60-DE (Double-junction pH/conductivity
probe), AI1201 PH60-E (Regular pH glass bulb probe), AI3711-E PH60-DE (Double-junction pH glass
bulb probe), AI1205 PH60S-E (Spear pH probe for solids/semi-solids pH testing), AI1203 PH60F-E (Flat
pH probe for surface pH testing), AI1209 EC60-E (Conductivity probe)
13
Battery Replacement
Please install batteries according to the following steps. *Please note the correct direction of battery installation:
The Positive Side (“+”) OF EVERY SINGLE Battery MUST FACE UP.
(WRONG INSTALLATION OF BATTERIES WILL CAUSE DAMAGE TO THE TESTER AND POTENTIAL HAZARDS!)