Version:1.0 User Manual Blood Pressure Monitor LC-BP01 Arm Type Guangdong Transtek Medical Electronics Co., Ltd. Zone B, No.105, Dongli Road, Torch Development District, Zhongshan, 528437, Guangdong, China Thank you very much for selecting moon Blood Pressure Monitor LC-BP01. Please read the user manual carefully and thoroughly so as to ensure the safe usage of this product, Keep the manual well for your further reference in case you have problems.
TABLE OF CONTENTS Table of Contents INTRODUCTION ................................................................ 2 General Description Indications for Use Contraindications Measurement Principle Safety Information LCD Display Signal Monitor Components List BEFORE YOU START ........................................................... 9 Power Supply and Charge Power Setting the Time,Date and Unit Pair up with Your Device Select the User Turn on or turn off the voice Tie the Cuff .............................
INTRODUCTION General Description Thank you for selecting moon arm type blood pressure monitor (LC-BP01). The monitor features blood pressure measurement, pulse rate measurement and the result storage. The design provides you with two years of reliable service. INTRODUCTION Safety Information The signs below might be in the user manual, labeling or other components. They are the requirement of standard and using.
INTRODUCTION INTRODUCTION CAUTION * This device is intended for adult use in homes only. * The device is not suitable for use on neonatal patients, pregnant women,patients with implanted, electronical devices, patients with pre-eclampsia, premature ventricular beats, atrial fibrillation, peripheral, arterial disease and patients undergoing intravascular therapy or arterio-venous shunt or people who received a mastectomy. Please consult your doctor prior to using the unit if you suffer from illnesses.
INTRODUCTION INTRODUCTION CAUTION * The service life of the cuff may vary by the frequency of washing, skin condition, and storage state. The typical service life is 10000 times. * It is recommended that the performance should be checked every 2 years and after maintenance and repair, by retesting at least the requirements in limits of the error of the cuff pressure indication and air leakage (testing at least at 50mmHg and 200mmHg).
BEFORE YOU START INTRODUCTION Monitor Components DC POWER SOCKET VOICE ROADCAST BUTTON Power Supply and Charge Power AIR CONNECTOR PLUG LCD DISPLAY 1. The battery of LC-BP01 is built-in rechargeable li-polymer battery, the battery current is 1000 mAh. 2.
BEFORE YOU START CAUTION • Do not attempt to replace your blood pressure monitor’s battery. It is built-in and not changeable. • Only charge the battery in accordance with the user instructions supplied with the blood pressure monitor. • Avoid charging your blood pressure monitor in extremely high or low temperatures. • Do not use your blood pressure monitor while you are charging it. • Do not attempt to disassemble the blood pressure monitor or force open the built-in battery.
BEFORE YOU START BEFORE YOU START Tie the Cuff Select the User Press and hold the on/off button for 3 seconds when the device is in the state of power-on to select the user . SYMBOL DESCRIPTION EXPLANATION User ID appears when the monitor is operated by User 1. User ID appears when the monitor is operated by User 2. User ID appears when the monitor is operated by guest .
MEASUREMENT MEASUREMENT Start Measurement Inflating and measuring. 1.When the monitor is off, press ON/OFF BUTTON to turn on the monitor. (When the volume is on,it will speak “ Starting measurement, please relax” first, and then complete the measurement process.) Boot animation Display and save the results.(When the volume is on, It will broadcast your measuring result.) Before your start 2.Press “Start/Stop” button to power off, otherwise it will turn off within 1 minute.
INFORMATION FOR USER INFORMATION FOR USER Tips for Measurement Maintenance Measurements may be inaccurate if taken in the following circumstances. To obtain the best performance, please follow instructions below. Within 1 hour after dinner or drinking Immediate measurement after tea, coffee, smoking Put in a dry place and avoid the sunshine Avoid intense shaking and collisions.
ABOUT BLOOD PRESSURE ABOUT BLOOD PRESSURE What are systolic pressure and diastolic pressure? When ventricles contract and pump blood out of the heart, the blood pressure reaches its maximum value in the cycle, which is called systolic pressure. When the ventricles relax, the blood pressure reaches its minimum value in the cycle, which is called diastolic pressure.
SPECIFICATIONS TROUBLESHOOTING If any abnormality arises during use, please check the following points: PROBLEM SYMPTOM No power Display will not light up. Unable to charge "Low power, please charge in time" is displayed. Error message CHECK THIS Power is exhausted. Click the ON/OFF button, please wait for 4 seconds to turn on the device Power is low. REMEDY Charge the power "Fault Code E02" The monitor detected is displayed . motion,talking or the pulse is too poor while measuring.
COMPLIED STANDARDS LIST ATHORIZED COMPONENT Athorized Component Please use the moon authorized adapter Adapter Type: BLJ06L050100U-U Input: 100-240V, 50-60Hz, 0.2A max 1000mA Output: 5V Contact Information For more information about our products, please visit www.moonolife.com You can get customer service, usual problems and customer download, Transtek will serve you anytime. Manufactured by: Guangdong Transtek Medical Electronics Co., Ltd. Company: Guangdong Transtek Medical Electronics Co., Ltd.
EMC GUIDANCE EMC GUIDANCE EMC Guidance Table 2 The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare environments. Warning: Don’t be near the active HF surgical equipment and the RF shielded room of an ME system for magnetic resonance imaging, where the intensity of EM disturbances is high. Warning: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation.
SPECIFICATIONS FCC STATEMENT EMC GUIDANCE FCC Statement Table 3 FCC ID:OU9LC-BP01-B Guidance and manufacturer’s declaration - electromagnetic Immunity Radiated RF Test Band Service IEC61000-4-3 Frequency (MHz) (MHz) (Test specifications 385 380-390 TETRA for ENCLOSURE 400 PORT IMMUNITY to 450 430-470 GMRS 460, RF wireless FRS 460 communications 710 704-787 LTE Band equipment) 13, 745 17 780 810 800-960 870 930 1720 1845 17001990 1970 Modulation Distance IMMUNITY TEST (W) (m) LEVEL (V/m) Pulse mod