Instruction manual
22
DIGITAL CALIBRATION
I.UtilizingtheDesolderingGunDigitalTemperatureCalibration
1. TurnontheDesolderinggunfunctionswitch.
2. Set to appropriate temperatureyou wantto calibrate. Place the
tipofthedesolderinggunonanexternaltemperaturemeter.
3. Thereadingsontheexternaltemperaturesensorshouldbemore
orlessequaltothedisplayedtemperature.
4. Iftherearelargediscrepancyinthetemperaturereadingwecan
recalibrate the temperature setting. First write down the set
temperature of the soldering iron and the actual temperature
readingfromtheexternaltemperaturemeter.
Forexample:
Settemperature=300
externaltemperature=350
5. TurnofftheSolderingIronFunctionswitch(“14”fromthecontrol
panel).EnsurethattheDesolderingGunFunctionswitchisinthe
offposition(“12”fromthecontrolpanel).
6. PressandholdtheUP buttonoftheSolderingIronTemperature
Adjustmentbutton(“4”fromthecontrolpanel).
7. TheDesolderingGunTemperatureDisplay (“14”fromthecontrol
panel).Willswitchtofourzeros“000”indicatingitisnowinthe
desoldering gun temperature calibration mode. Release the UP
buttonoftheSolderingIronTemperatureAdjustmentbutton(“4”
fromthecontrolpanel).
8. Use the Desoldering Gun Temperature Adjustment buttons (“5”
from the control panel) to increase or decrease the calibration
values.Inourexamplethesettemperatureis300buttheactual
temperature is 350, There is a difference of 50 degrees. Press
theupbuttonuntilwereach“050”.
9. Confirmthe changebypressingandholdingthe“UP” buttonof
the SolderingIronTemperatureAdjustmentbutton(“4”fromthe
controlpanel).
11
OPERATING GUIDELINES
4. Adjust the air flow level using the AIR PRESSURE ADJUSTMENT
BUTTONS(“3”fromthecontrolpanel).
5. Adjust the hot air gun air temperature using the HOT AIR GUN
TEMPERATURE ADJUSTMENT BUTTONS (“2” from the control
panel). The prefix of the display for Hot Air Gun Temperature will
change from “C” to “c” indicating that air temperature is being
adjusted. It will return to “C” indicating actual temperature is
displayed.
6. Reworking task can be started 1 minute after the desired hot air
temperature and airflow level are reached, as also indicated from
displaypanel“18”.
7. When reworking is complete, return the Hot Air Gun to its holder
butDONOTimmediatelyturnoffthemainpowerswitch.
8. DeactivatetheHotAirGunActivationbuttonfirstinordertoactivate
theautocoolingprocess.Thesystemwillstarttoblowair(atroom
temperature)atafastratetoreduceheatfromthehotairgunand
bring down the temperature to a reasonable safe level of 90°C.
Duringthistime,theairflowdisplaywillalsochange“cool”while
temperatureisgraduallydecreasing.Oncethetemperaturedropsto
approximately 90°C thesystemwillhaltanddisplay“OFF”onthe
panel.Itisnowsafetoswitchoffthemainpowerswitch.
9. TurnOFFthemainpowerswitch.
10. Unplugthedevicefromthemainpowersource.
IMPORTANT:Whenadjustingtheairtempe rature,itisstrongly
advisedtoinitiallyincrease theairflowlevelinorderto manage
the system temperature. This is to protect the heating element
insidethehandlefromexcessiveheatandavoidthepossibilityof
subjectingadjacentcomponentstothermalshock.