IMPORTANTREMINDERS:
1. Make sure the equipment is placed on a flat stable surface
and all the heatgenerating components placed on their
respectiveholdersorstands.
2. EnsureallfunctionswitchesareOFFpriortoreworking.
3. Ensureallterminalconnectionsareproperlysecured.
IMPORTANT: Please refer to the CONTROL PANEL GUIDE
page for buttons and display panel directory.
OPERATING GUIDELINES
A.INITIALPROCEDURES
1. Plug the device to the main power source using the power cord
providedinthepackage.
2. With all function switches deactivated and all terminal connections
properlysecured,switchONthedevicebyactivatingthemainpower
switch(“10”fromthecontrolpanel).
3. The display panels, will temporarily show the product name “AOYUE
Int2702”andthendisplay“OFF”.Thesystemwillremainatthisstate
untiltheuseractivatesafunction.
B.HOTAIRGUN
1. Followtheinitialproceduresabove,“A.INITIALPROCEDURES”.
2. Activate“HotAirGun”switch(“19”fromthecontrolpanel).
3. Thesystemwillimmediatelystarttoblowairatanairflowrateof50
while rapidly and safely increasing the air temperature to 100°C
(defaultsystemoperatingparameters).Thesevalueswillbereflected
from the Hot Air Gun Air Temperature and Air Pressure display
panels,“18”and“16”fromthecontrolpanel,respectively.Themetal
ball inside the air gauge (“1” from the control panel) will also settle
somewhereinthemiddleofthevisibleareaindicatingthatthesystem
isblowingairatanaverageornormalrate.
23
DIGITAL CALIBRATION
SolderIronDigitalTemperatureCalibrationExample
● Theexternaltemperaturesensordisplays250degrees.
● The set temperature and displayed actual temperature of the
solderingironis300degrees.
● 300 — 250 = 50. An additional adjustment of 50 degrees is
required.Uponenteringcalibrationmode,thedisplayshows“010”,
indicatingacalibrationnumberof10isalreadypresent.
● Therefore10+50=60.
● Weadjust from“010”to “060”by pressingtheupbuttonofthe
Soldering Iron Temperature Adjustment button (“4” from the
controlpanel)
● Saveandexitcalibrationmode.
● Theexternaltemperaturesensorwouldnowdisplay298to302.
DesolderingGunDigitalTemperatureCalibrationExample
● Theexternaltemperaturesensordisplays300degrees.
● The set temperature and displayed actual temperature of the
solderingironis350degrees.
● 300 — 350 = 50. An additional adjustment of –50 degrees is
required.Uponenteringcalibrationmode,thedisplayshows“010”,
indicatingacalibrationnumberof10isalreadypresent.
● Therefore1050=50.
● Weadjustfrom“010”to“050”bypressingthedownbuttonofthe
Desoldering Gun Temperature Adjustment button (“5” from the
controlpanel)
● Saveandexitcalibrationmode.
● Theexternaltemperaturesensorwouldnowdisplay298to302.
NOTES:
● Calibration will only make the newly calibrated point the most
accurate.Othertemperaturepointsmaybealittleoff.