Instruction manual

10
IMPORTANTREMINDERS:
1. Make sure the equipment is placed on a flat stable surface
and all the heatgenerating components placed on their
respectiveholdersorstands.
2. EnsureallfunctionswitchesareOFFpriortoreworking.
3. Ensureallterminalconnectionsareproperlysecured.
IMPORTANT: Please refer to the CONTROL PANEL GUIDE
page for buttons and display panel directory.
OPERATING GUIDELINES
A.INITIALPROCEDURES
1. Plug the device to the main power source using the power cord
providedinthepackage.
2. With all function switches deactivated and all terminal connections
properlysecured,switchONthedevicebyactivatingthemainpower
switch(“10”fromthecontrolpanel).
3. The display panels, will temporarily show the product name “AOYUE
Int2702”andthendisplay“OFF”.Thesystemwillremainatthisstate
untiltheuseractivatesafunction.
B.HOTAIRGUN
1. Followtheinitialproceduresabove,“A.INITIALPROCEDURES.
2. Activate“HotAirGun”switch(“19”fromthecontrolpanel).
3. Thesystemwillimmediatelystarttoblowairatanairflowrateof50
while rapidly and safely increasing the air temperature to 100°C
(defaultsystemoperatingparameters).Thesevalueswillbereflected
from the Hot Air Gun Air Temperature and Air Pressure display
panels,“18”and“16”fromthecontrolpanel,respectively.Themetal
ball inside the air gauge (“1” from the control panel) will also settle
somewhereinthemiddleofthevisibleareaindicatingthatthesystem
isblowingairatanaverageornormalrate.
23
DIGITAL CALIBRATION
SolderIronDigitalTemperatureCalibrationExample
Theexternaltemperaturesensordisplays250degrees.
The set temperature and displayed actual temperature of the
solderingironis300degrees.
300 — 250 = 50. An additional adjustment of 50 degrees is
required.Uponenteringcalibrationmode,thedisplayshows“010”,
indicatingacalibrationnumberof10isalreadypresent.
Therefore10+50=60.
Weadjust from“010”to “060”by pressingtheupbuttonofthe
Soldering Iron Temperature Adjustment button (“4” from the
controlpanel)
Saveandexitcalibrationmode.
Theexternaltemperaturesensorwouldnowdisplay298to302.
DesolderingGunDigitalTemperatureCalibrationExample
Theexternaltemperaturesensordisplays300degrees.
The set temperature and displayed actual temperature of the
solderingironis350degrees.
300 — 350 = 50. An additional adjustment of 50 degrees is
required.Uponenteringcalibrationmode,thedisplayshows“010”,
indicatingacalibrationnumberof10isalreadypresent.
Therefore1050=50.
Weadjustfrom“010”to“050”bypressingthedownbuttonofthe
Desoldering Gun Temperature Adjustment button (“5” from the
controlpanel)
Saveandexitcalibrationmode.
Theexternaltemperaturesensorwouldnowdisplay298to302.
NOTES:
Calibration will only make the newly calibrated point the most
accurate.Othertemperaturepointsmaybealittleoff.