User Manual
PEN TYPE DIGITAL MULTIMETER
OPERATOR’S MANUAL
1. Overview
The pen type digital multimeter is characterized at slim size,
portable, stable performance and anti-dropping capacity. Using
2000 counts digit LCD monitor with character 12mm high, they
offer clear readings. With overall circuitry design centering on
large-scale IC A/D converters in conjunction and over-load
protection circuit, the meters give excellent performance and
exquisite making as a handy utility instrument.
The meters can be used to measure DC & AC voltage, resistance,
positive diode voltage fall and audible continuity. If the equipment
is used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired.
2. Panel Layout
1. VΩ: V/ Input test lead.
2. Lamp light.
3. Test lead fixture: Fix the COM test lead.
4. LCD display: 3 1/2 digits LCD display, full function symbol
display.
5. Toggle Switch: Use this switch to select functions and ranges.
6. HOLD key: Press the “HOLD” key to lock display value, and the
“DH” sign will appear on the display, press it again to exit.
7. Key: Press the “ ” key over 2 seconds to light the back light,
press it again for more than 2 seconds to lights off it, it can lights
off automatically after 15 seconds too.
8. SELECT key: This key work on the " " range, press the
key to choose resistance, diode or continuity test, on the voltage
range, change to DC or AC.
9. Protective casing.
10. Battery cover.
11. COM: COM input test lead.
3. Safety Information
3-1 The meter is designed according to IEC-1010 concerning
electronic measuring instruments with an over-voltage category
600V (CAT Ⅲ) and pollution 2.
3-2 Follow all safety and operating instructions to ensure that the
meter is used safely and is kept in good operating condition.
3-3 safety symbols:
Important safety information, refer to the operating manual.
Dangerous voltage may be presence.
Double insulation (protection Class II)
4. Special Cautions for Operation
4-1 To avid risk of electric shock, do not use the meters before the
cover is in place.
4-2 The toggle switch should be right position for the testing.
4-3 To avoid electric shock and damaging the instruments, the
input signals are forbidden to exceed the specified limits.
4-4 When measuring TV set or switched power, attention should
be paid to the possible pulses that may bring destruction to the
circuit.
4-5 Toggle switch position is forbidden to be changed at random
during measurement.
4-6 Take caution against shock in the course of measuring voltage
higher than DC 60V & AC 30V.
4-7 After operation is finished, set toggle switch at OFF to save
battery power.
4-8 If the meter is without usage for long time, take out battery to
avoid damage by battery leakage.
5. GENERAL SPECIFICATIONS
5-1 Max Voltage between input terminal and Earth Ground: CAT Ⅲ
600V.
5-2 Over-range Indication: display “OL” for the significant digit.
5-3 Automatic display of negative polarity “-”.
5-4 Low Battery Indication: “ ” displayed.
5-5 Display: 3 1/2 digit LCD with a max. reading of 1999.
5-6 Auto range control.
5-7 Auto Power Off: When measurement exceeds 15 minutes
without switching mode and pressing key, the meter will switch
to standby mode. Press any key to exit standby mode.
5-8 Power supply: 1.5V×2 “AAA” R03P battery
5-9 Operating Temp.: 0℃ to 40℃ (relative humidity <85%)
5-10 Storage Temp.: -10℃ to 50℃ (relative humidity <85%)
5-11 Guaranteed precision Temp.: 23±5℃
(relative humidity <70%)
5-12 Dimension: 235 x 54 x 30mm
5-13 Weight: approx. 200g (including battery)
6. Testing Specifications
Accuracy is specified for a period of year after calibration and at
18℃ to 28℃ (64℉ to 82℉) with relative humidity to 70%.
6-1 DC Voltage
Range
Resolution
Accuracy
200mV
0.1mV
±(0.5% of rdg + 2 digits)
2V
1mV
20V
10mV
200V
100mV
600V
1V
±(0.8% of rdg + 2 digits)
-- Impedance: 10MΩ, More than 100MΩ on 200mV range
-- Overload protection: 600V DC or AC rms
6-2 AC Voltage
Range
Resolution
Accuracy
2V
1mV
±(0.8% of rdg + 3 digits)
20V
10mV
200V
100mV
600V
1V
±(1.2% of rdg + 3 digits)
-- Impedance: 10MΩ
-- Overload protection: 600V DC or AC rms
-- Frequency Range: 40 to 400Hz
-- Response: average, calibrated in rms of sine wave
6-3 Resistance
Range
Resolution
Accuracy
200Ω
0.1Ω
±(1.0% of rdg + 3 digits)
2kΩ
1Ω
±(1.0% of rdg + 2 digits)
20kΩ
10Ω
200kΩ
100Ω
2MΩ
1kΩ
20MΩ
10kΩ
±(1.5% of rdg + 3 digits)
-- Overload protection: 600V DC or AC rms
6-4 Diode and Audible continuity test
Range
Description
Test Condition
Display read
approximately
forward voltage of
diode
Forward DC current
approx. 0.4mA
Reversed DC voltage
approx. 1.5V
Built-in buzzer
sounds if resistance
is less than 50Ω
Open circuit voltage
approx. 0.5V
-- Overload protection: 600V DC or AC rms
7. OPERATING INSTRUCTIONS
7-1 Attention before operation